Users Manual

Page 13
Specification
Quattro Vanguard
Solo dB
Size (W x D x H)
16.2 x 20.6 x 35.8 in (411 x 524 x 908 mm)
Capacity (pans/bottles)
1
Capacity (wash bowls)
1
Cistern capacity
3 US gallons (11 litres)
Inlet protected by ‘EN 13077, Family A, Type B’
Electrical supply
120 v / 1 Ph / 60 Hz
Energy use
0.01 KWh / cycle
Footprint
2.31 ft
2
(0.215 m
2
)
Height (lid open)
47.97 in (1218 mm)
Humidity operating range
50-80% RH
Loading height
30.87 in (784 mm)
Motor power
600 w
Pump power
125 w
Temperature operating range
+41 to +104°F (+5 to +40°C)
Water requirements
Cold 1.5 US gallons/min (5.5 lpm)
Weight (dry)
123 Ibs (56 kg)
Open / close
Foot Pedal open and close
Auto-start
Maintenance access
All sides
Cistern overflow indicator
Never connect the overflow indicator stub to the drain.
Ensure any discharge is visible
Direction of rotation for macerator
motor
Single phase machines are supplied with the correct rotation.
Isolator
For single and three phase machines, 20A isolator to National Electricity
Code standards
Mains water inlet flow rate
Required flow rate is minimum of 1.5 US gallons/min (5.5 litres/min)
Water regulated to 2.4 US gallons/min (9 litres/min ) maximum
Mounting
Floor. 3/8” dia securing bolts (10mm dia securing bolts)
Pipework (mains)
Inlet 3/4GHT female connection
Pipework (waste drain outlet)
50mm (2” HT) multifit P trap
Water pump
Self-primed from the cistern. Electric centrifugal pump
All installations must comply with statutory regulations, local water by-laws and relevant codes of practice of the
country of installation. Responsibility for this must rest with the installer. Haigh make every effort to comply with
national requirements/standards.