Users Manual

Page 30
Cleaning recommendations
Electrically isolate the machine before cleaning
Never use a wet solution to clean the indicator panel
Daily
All exterior panels to be wiped over with normal cleaning solution for worktops etc. (soapy water) and then
dried.
The best results are obtained by opening the lid which allows full access to the seating and the lid seal.
All internal surfaces are automatically cleaned by the machine. Failure to do so must be investigated.
Weekly (as required)
The lid seating area should be scrubbed with a brush, wiped and the neoprene seal washed with the same
cleaning solution.
Lubrication
The machine is designed for minimum maintenance.
The clip bushes used in this machine must not be lubricated.
Apply anti-seize compound if indicated.
The motor is fitted with sealed for life bearings.
The mechanical seal face must be perfectly clean.
Use only clean water to lubricate the seal face.
Ordering spare parts
Spares parts are identified on pages 34 to 45.
Please contact your local distributor for more information.
Please quote the following information:
Your contact details
The machine serial number
The part number required
The full part description
The quantity of each part required
The invoice address
The delivery address.
Many parts are available within 24 hours. We can provide technical advice should you need it.