13Operating Instructions VisiTrace
mA Sensors
INSTALLATIONPRODUCT DESCRIPTION
Full online wireless option via Bluetooth 5 for easy
monitoring, configuration and calibration
Encrypted data exchange with mobiles (Android & iOS)
4.2 Hardware Description
The VisiTrace mA sensor consists of a sensor head with inte-
grated electronic and a sensor shaft in contact with the measured
medium. The sensor shaft is terminated by the optical dissolved
oxygen (ODO) cap, carrying the oxygen sensitive luminophore.
During development, special attention was paid to an optimum
sanitary design.
Sensor status LED of the sensor:
LED Status Case
All LEDs light up shortly
one by one in a circle Power Up
Geen LEDs are flashing Ready to operate, no errors or
warnings have been registered
Yellow LEDs are flashing Warning, minimum one
warning is active
Red LEDs are flashing Error, minimum one error is
active
All LEDs light up slowly Wireless communication to
one by one in a circle ArcAir is active
Red LED / Yellow LED / Green LED
M12 Connector
Sensor Head with
Transmitter including
Bluetooth antenna
Process connection
PG13,5 stainless steel
O-ring
Sensor shaft with stainless
steel heat number and REF
ODO Cap with oxygen-sensitive
luminophore (sensing element), ODO
Cap serial number, Ref number and
stainless steel heat number
Protection Cap
ODO Cap L0
REF
243461/0 0
LOT
1110 669 55
SN
114 3
2019-05-07
Figure 4: VisiTrace mA description
4.3 Optical DO measurement
The optical measurement principle is based on the so-called
luminescence quenching. The luminescence of certain organic
pigments (luminophore) is quenched in the presence of oxygen.
The luminophore absorbs the excitation light and release a part of
the absorbed energy by emission of fluorescence. In the presence
of oxygen, energy transfer takes place from the excited lumino-
phore to oxygen. The luminophore does not emit fluorecence and
the measurable fluorescence signal decreases.
O
2
O
2
1
2
Light (excitation)
Light Absorption Excited State
Luminophore
Fluorescence process (without oxygen)
Quenching in the presence of oxygen
Energy transfer
Light Absorption Excited State
Fluorescence
(emission)
No fluorescence
Figure 5: Fluorescence quenching by oxygen
4.4 VisiTrace mA with Micro-Transmitter
inside
With the micro-transmitter integrated, VisiTrace mA sensors offer
fully compensated signal directly to the process control system.
Communication protocols include standard analog 4–20 mA or
digital HART. The micro-transmitter located in the sensor head
stores all relevant sensor data, including calibration and diag-
nostic information, simplifying calibration and maintenance.
5 Installation
5.1 Unpacking
1) Unpack carefully the VisiTrace mA sensor. Enclosed you
will find the VisiTrace mA sensor, the Declaration of
Quality, the VisiTrace mA Operating Instructions, and the
Stainless Steel Inspection Certificate.
2) Inspect the sensor for shipping damages or missing parts.