User Manual
4
The switching capacity is not more than 24V / 3A.
When the access control needs to be output, connect the access control signal wire
to the terminal block.
A) when only the wrist strap is tested and the test passes, the access control output.
C) when only two foot shoes are tested and the test passes, the access control
output.
D) when testing the wrist strap and feet shoes at the same time, when the test
passes, the access control output.
Operation
1. Please plug the power adapter into the power socket, and the static DC current is
9V / 40mA.
2.ESD test, as long as you step on the opposite metal plate with your feet, wear a
good hand and brain strap to touch the wiring port, and gently press the metal test
button with your fingers, you can know whether the ESD equipment is qualified at a
glance.
Note: do not wear gloves or press metal test button with fingernail
Installation
1. Thread the wiring on the pedal from the bottom of the bracket to the top.
2. Three sets of m8x25mm to fix bracket and pedal.
3. Broadcast the wiring to the interface at the bottom of the instrument.
4. Align the three holes at the bottom of the instrument with the three screw posts
on the bracket base, and then screw them down to fix them.
5. Tighten the lifting lock of the support and lock the height of the support.