Phase 0
H.110: the counter direction
change.
U3.0.36
Grinding Direction of
Phase 1
H.010: default direction
H.110: the counter direction
H.110
If you change it to
be H.010, it will
be single direction
grinding.
U3.2.11
Timing Control
4200: untimed
4239: timing
4239
If timing grinding
is not needed,
modify it.
U3.2.17
Paused Interval Time
0.0~3600.0s
0000.0
Extend or shorten
the time.
U3.2.24
Total Grinding Time
0.0~3600.0min
0600.0
U5.0.15
Displayed Ratio
0.0001~6.5000
0.1600
Always keep it!
U5.0.19
Restore Factory
Setting
000: no operation
030: backup the data
060: recovery the data
102: restore factory settings
000
Modify it upon
your actual needs.
4.7.4 Setting the Grinding Parameters
Example: modify parameter U5.0.19 to ‘backup the data’
Steps
Buttons
Display
Specifications
1
Connect with the power
Ball mill display mode.
2
Press [MODE]
Start the mode modification.
3
Press [] 6 times
Find the parameter U5.0.19.
4
Press [ENTER]
The parameter of U5.0.19 can be modified.
5
Press [>>] 2 times
Choose the second value.
6
Press [] 3 times
Value changes from 0 to 3.
7
Press [ENTER]
The modification of U5.0.19 is confirmed.
8
Press [MODE]
Back to the ball mill display mode.
4.8 Performing a Grinding Operation
After everything has been set up as described in section 4 Working with the mill, close
the hood.