User Guide

13
12
C 102
Range Turbidity 0.00 to 50.0 NTU*
Bromine 0.00 to 8.00 mg/L
Free Cl 0.00 to 2.50 mg/L
Total Cl 0.00 to 3.50 mg/L
CYS 0 to 80 mg/L
Iodine 0.0 to 12.5 mg/L
LR Iron 0.00 to 1.00 mg/L
pH 5.9 to 8.5
Resolution
Turbidity 0.01 and 0.1 NTU*
Bromine 0.01 mg/L
Free Cl 0.01 mg/L
Total Cl 0.01 mg/L
CYS 1 mg/L
Iodine 0.1 mg/L
LR Iron 0.01 mg/L
pH 0.1
Accuracy
Turbidity ±0.5 NTU* or ±5% (whichever greater)
Bromine ±0.08 mg/L; ±3%
Free Cl ±0.03 mg/L; ±3%
Total Cl ±0.03 mg/L; ±3%
CYS ±1 mg/L; ±15%
Iodine ±0.1 mg/L; ±5%
LR Iron ±0.02 mg/L; ±3%
pH 0.1 pH
Light Source Pure Green LED
Light Source Life Life of the instrument
Light Detector Two Silicon Photocells
Power Source 4 x 1.5V AA alkaline batteries
Battery Life 60 hours or 1000 measurements
Automatic Selectable
Shutdown 10, 20, 30, 40, 50 or 60 min.
Operating 0 to 50°C (32 to 122°F)
Conditions 95% RH max (non-condensing)
Dimensions 220 x 82 x 66 mm
(8.7 x 3.2 x 2.6")
Weight 510 g (18 oz.)
SPECIFICATIONS
* 1 NTU = 1 FTU
shown. If pressed in
time/date setup mode, the
meter will quit current
mode without making any
changes to current time/
date.
12) key Scrolls upwards through
the parameters to be mea-
sured. In calibration/diag-
nostic mode, increments
the blinking digit by one.
If pressed together with
ALT while the meter is in
logging mode, the upper
LCD will show the data
(date/time/value) in the
memory.
13) ALT key: Alternative functions.
14) key: Scrolls downwards through
the parameters to be mea-
sured. In calibration/diag-
nostic mode, decreases
the blinking digit by one.
If pressed together with
ALT while the meter is in
logging mode, the upper
LCD will show the current
lot number.