CALIBRATION
TURBIDITY CALIBRATION
The meter should be properly calibrated with
a standard prepared as described in USEPA
method 180.1.
Span calibration
To calibrate the span of the meter,
fill the cuvet with the primary turbid-
ity formazine standard of 50 NTU.
Inspect and clean thoroughly the sur-
face of the vial.
Shake the standard vigorously for a
few seconds and wait a few minutes
for the bubbles to disappear.
Turn the meter on and press both ALT and
CAL momentarily.
The LCD will show four
dashes on the upper
and d 00 on the lower
part of the LCD with
the second 0 blinking.
Using the , and
keys change the lower
row of the display to
show d 21.
Insert the previously prepared 50 NTU stan-
dard into the cuvet holder and make sure that
the notch on the cap is positioned securely
into the groove. Press .
+
positioned securely into the groove. Press
READ.
The meter will first show SIP for a few
seconds and afterwards the sample concen-
tration: