User Guide

29
28
of the display to remind the user that every
time a measurement is taken, the value is stored
in the next available lot number.
REVIEWING THE LOGGED BUFFER
To review the memo-
rized values, press to-
gether ALT and .
The meter will scroll all the data in the buffer
showing the lot number, value, date and time.
e.g. The first recorded reading in the buffer is
lot # 0, 0.35 mg/L of Free Chlorine,
memorized on 23
rd
August at 3:34 pm;
The second logged data relates to lot
# 1, 1.35 mg/L of the customized param-
eter, logged on 23
rd
August at 3:55 pm.
First lot Second lot
+
CLEARING THE MEMORY
After all the buffer (memory) is taken up, the
LCD will blink Full.
LOGGING WITH C 114
C 114 allows user to log 25 time/day-tagged
measurements. User can easily turn the log-
ging mode on and off, review the logged
memory, review the current lot number and
clean the buffer (memory). C 114 also re-
minds user if its memory is full.
TURNING THE LOG MODE ON OR OFF
Enter the diagnostic
mode by pressing ALT
and CAL together.
Select mode 12 and press the
key.
The display will show the current (vacant) lot
together with LOG if the log mode is on.
Otherwise it will show ---- if the log mode is
off.
By pressing the key, the meter toggles
between the Log on and off positions. If the
log on mode is selected, every time a mea-
surement is taken (READ pressed) the rel-
evant values will be stored in the current
(vacant) lot number.
To quit diagnostic mode,
press the ALT and CAL
keys together again.
The LCD will then show CAL and Stor for a
few seconds. If the log on mode was selected
LOG will appear on the bottom left hand side
+
+
To quit diagnostic mode,
press the ALT and CAL