CALIBRATION
If repeated measurements differ more than an acceptable value, check if the meter
calibration status by measuring a standard solution.
Anyway, it is recommended to perform a monthly calibration of the meter, or more
frequently for greatest accuracy.
To check the date of last calibration, simply press the GLP/CAL key. Press again to
toggle between date and time.
CALIBRATION PROCEDURE
Turn the meter on and wait for the display to show
dashes “----”.
Press ALT+CAL keys, the “CAL” message will blink on the display 3 times, then
the meter enter the calibration mode asking for the 0 EBC standard.
Fill the measurement cuvette with C124-0 stan-
dard (@ 0 EBC).
Note: In order to minimize any error introduced by the cuvette, it is recommended
to use the same cuvette during calibration and for measurements.
Insert the cuvette with the 0 EBC standard solution (or turbidity-free dilution
water) into the measurement cell and press the CAL key. “SIP” and “CL” will
start blinking.
After approximately 30 seconds the instrument
will ask for the 2.5 EBC standard solution.
Repeat the same procedure with C124-1 standard solution (@ 2.5 EBC): fill
the cuvette the C124-1 standard, insert into the measurement cell and press
the CAL key again.