User Guide

10
CALIBRATION
If repeated measurements differ more than an acceptable value, check if the meter
calibration status by measuring a standard solution.
Anyway, it is recommended to perform a monthly calibration of the meter, or more
frequently for greatest accuracy.
To check the date of last calibration, press and hold the DATE/ key down for few
seconds.
CALIBRATION PROCEDURE
Turn the meter on and wait for the
display to show “----”.
Press the CAL key once, the “CAL”
message will blink on the display for
about 6 seconds, then the calibration
mode stops.
While the “CAL” message is still blinking, press CAL again. The instrument is
now in the calibration mode and a “CL” will appear on the lower part of the
display.
The date of calibration can be edited
now by simply pressing the DATE/
key. To scroll to the correct number press
the READ/ key.
The first blinking parameter is the month, on the left hand of the display
(MM.DD).
To confirm the displayed data values
and to go to the next step, press the
CAL key once. A blinking “ZERO” mes-
sage will appear.
Fill the measurement cuvette with HI 93703-0
standard (@ ZERO FTU, or turbidity-free water).
Note: In order to minimize any error introduced by the
cuvette, it is recommended to use the same cuvette during calibration and for
measurements.
Insert the cuvette with the ZERO FTU
standard solution (or turbidity-free
dilution water) into the measurement
cell and press the CAL key.