User Guide

24
The instrument directly displays concentration in mg/L of ammonia
nitrogen (NH
3
-N) on the Liquid Crystal Display.
To convert the reading to mg/L of ammonia (NH
3
), multiply by a
factor of 1.214.
INTERFERENCES
Interference may be caused by:
Organic compounds like: chloramines, various aliphatic and aromatic
amines, glycine or urea above 10 ppm N (positive error).
To eliminate these interferences distillation is required.
Organic compounds like: aldehydes, alcohols (e.g. ethanol) or
acetone above 0.1 % (negative error).
To eliminate these interferences distillation is required.
Sulfide: may cause turbidity.
Ammonia LR
57
Mark the vial with a pencil on the
white band to place an orientation
sign. Use this sign to insert the test
vial always in the same position into
the holder.
Place the vial into the holder and
push it completely down.
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the
display will show “-0.0-”. Now the
meter is zeroed and ready for
measurement.
Remove the vial.
Remove the cap and add the content
of one packet of HI 93758-0
Phosphorus Reagent.
Replace the cap tightly and shake
gently to mix for 2 minutes until
most of the powder is dissolved. This
is the reacted sample.
Place the vial into the holder and
push it completely down. Ensure that
the vial orientation respect to the
holder is the same as the blank.
Press TIMER and the display will show the countdown prior to the
measurement. Alternatively, wait for 3 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
Phosphorus, Reactive