Mark the vial with a pencil on the
white band to place an orientation
sign. Use this sign to insert the test
vial always in the same position into
the holder.
Place the vial into the holder and
push it completely down.
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the
display will show “-0.0-”. Now the
meter is zeroed and ready for
measurement.
Remove the vial.
Remove the cap and add 4 drops of
HI 93764-0 Nessler Reagent.
Replace the cap tightly and mix by
inverting the vial a couples of times.
This is the sample.
Place the vial into the holder and
push it completely down. Ensure that
the vial orientation respect to the
holder is the same as the blank.
Press TIMER and the display will show the countdown prior to the
measurement. Alternatively, wait for 3 minutes and 30 seconds
and press READ DIRECT. In both cases "SIP" will blink during
measurement.
Ammonia HR
55
COD HR
Select the program number corresponding to Oxygen Demand,
Chemical HR (COD) on the secondary LCD by pressing PROGRAM
t and s.
Place the blank vial into the holder
and push it completely down.
Press ZERO and "SIP" will blink on the display.
Wait for a few seconds and the
display will show “-0.0-”. Now the
meter is zeroed and ready for
measurement.
Remove the blank vial.
Place the sample vial into the holder
and push it completely down.
Press READ DIRECT and "SIP" will blink during measurement.
Multiply the reading on the Liquid Crystal Display
by 10 to
obtain the concentration in mg/L of oxygen demand.
INTERFERENCES
Interference may be caused by:
Chlorides (Cl
-
): above 20000 mg/L.
Samples with higher chloride concentration should be
diluted.