User Guide

32
Place the vial into the holder and
push it completely down.
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the
display will show “-0.0-”. Now the
meter is zeroed and ready for
measurement.
Remove the vial.
Remove the cap and add the content
of one packet of HI 93711-0 Total
Chlorine Reagent.
Replace the cap and shake gently to
mix for about 20 seconds. This is
the reacted sample.
Place the vial into the holder and
push it completely down. Ensure that
the vial orientation respect to the
holder is the same as the blank.
Press TIMER and the display will show the countdown prior to the
measurement. Alternatively, wait for 2 minutes and 30 seconds
and press READ DIRECT. In both cases "SIP" will blink during
measurement.
The instrument directly displays concentration in mg/L of total
chlorine on the Liquid Crystal Display.
Chlorine, Total
49
Select the program number corresponding to Oxygen Demand,
Chemical LR (COD) on the secondary LCD by pressing PROGRAM
t and s.
Place the blank vial into the holder
and push it completely down.
Press ZERO and "SIP" will blink on the display.
Wait for a few seconds and the
display will show “-0.0-”. Now the
meter is zeroed and ready for
measurement.
Remove the blank vial.
Place the sample vial into the holder
and push it completely down.
Press READ DIRECT and "SIP" will blink during measurement.
The instrument directly displays concentration in mg/L of oxygen
demand on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Chlorides (Cl
-
): above 2000 mg/L.
Samples with higher chloride concentration should be
diluted.
COD LR