User Guide

11
Whenever the cuvet is placed into
the measurement cell, it must be
dry outside, and completely free of
fingerprints, oil or dirt. Wipe it
thoroughly with HI 731318 or a
lint-free cloth prior to insertion.
It is possible to take multiple
readings in a row, but it is
recommended to take a new zero
reading for each sample and to use
the same cuvet for zeroing and
measurement.
It is important to discard the sample immediately after the reading
is taken because the glass might become permanently stained.
All the reaction times reported in this manual are referred to 20°C
(68°F). As a general rule of thumb, they should be doubled at
10°C (50°F) and halved at 30°C (86°F).
42
Press ZERO and “SIP” will appear on the display.
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
Remove the cuvet and add
the content of one packet
of HI 93750B reagent. Re-
place the cap and mix
gently for one minute by
turning slowly the cuvet up-
side down.
Reinsert the cuvet into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and press READ
DIRECT. In both cases “SIP” will blink during measurement.
The instrument directly displays concentration in mg/L (ppm) of
potassium (K).
To convert the reading to mg/L of potassium oxide (K
2
O), multiply
by a factor of 1.20.
INTERFERENCES
The following ions do not interfere below the concentrations reported:
Ammonium 10 ppm
Calcium 10000 ppm as CaCO
3
Chloride 12000 ppm
Magnesium 8000 ppm as CaCO
3
Sodium 8000 ppm
Potassium MR
HI 83215 - NUTRIENT ANALYSES
Code Parameter Page
1 Ammonia LR 21
2 Ammonia MR 23
3 Ammonia HR 25
4 Nitrate LR 27
5 Nitrate MR 29
6 Nitrate HR 31
Code Parameter Page
7 Phosphorus LR 33
8 Phosphorus MR 35
9 Phosphorus HR 37
10 Potassium LR 39
11 Potassium MR 41
12 Potassium HR 43
PARAMETERS REFERENCE TABLE