User's Manual

32
HI 3002 FOR IN-LINE INSTALLATION
The drawing illustrates the ideal installation system since the stream
pressure in the pipe forces the air bubbles out automatically.
HI 3001D illustrated preaviously can also be mounted in this fashion.
However the longer
stem of HI 3002 fa-
cilitates an easier
escape route for any
air bubbles. HI 3002
comes with external
1/2” NPT and
thread 60 mm (2.4”)
immersion level.
HI 7638 FOR IN-LINE INSTALLATION
HI 7638 comes with
stainless steel external
3/8” NPT thread and
60 mm (2.4”) immer-
sion level. It can work
at temperatures of up
to 120 °C and should
be used in high tem-
perature/pressure applications such as in boilers (HI 9923).
HI 3002 FOR TANK INSTALLATION
Due to its Ultem® and sealed body, HI 3002 can be installed
directly in a tank or vat without requiring a probe holder.
> 1/2
GAS
HI 3002
3/4
GAS
3/4to 3/8
nipple
HI 7638