User's Manual

22
Press the CAL key. The temperature compensated
buffer value will be displayed e.g. if the buffer
temperature is at 25°C, the display will show
pH7.01. If the buffer temperature is at 20°C,
display will show pH 7.03 (see page 23). If
"E4" appears the calibration solution is wrong.
Wait until the pH symbol stops flashing: that
means the reading is stable.
Press the CON key. The symbol "E5" will appear
to indicate that the electrode is still in pH 7.01
buffer, but expecting a pH 4.01 or 10.01 buffer.
Rinse the electrode and temperature probe be-
fore immersing them into pH 4.01 / pH 10.01 calibration buffer,
shake briefly and wait for thermal equilibrium.
The symbol "E5" should disappear and the temperature compen-
sated buffer value is flashing.
Wait until the pH symbol stops flashing, press the CON key to
confirm the calibration.
The pH calibration is now complete.
CALIBRATION WITH MANUAL TEMPERATURE COMPEN-
SATION FOR HI8417, HI8520 AND HI8521 ONLY)
If for some reason the temperature probe is defective, or it is required
to calibrate with manual temperature compensation, follow the proce-
dure below:
Unplug the temperature probe from the
meter.
Note the temperature of the buffer with a
ChecktempC or a glass thermometer.
Use the UP and DOWN arrow keys to
set the temperature to the noted tem-
perature.
Follow the calibration procedure as for
pH calibration with temperature probe
connected (see page 18 for HI8417 or
page 21 for HI8520/HI8521).
°
C
°
C
ºC
CAL
CON
pH