Up to 100 logged samples can be stored into memory.
LOGGING THE CURRENT DATALOGGING THE CURRENT DATA
LOGGING THE CURRENT DATALOGGING THE CURRENT DATA
LOGGING THE CURRENT DATA
To store the current reading into memory press the LOG key while in
measuring mode.
The instrument will display the current date (mm.dd) on the primary
LCD, the record number on the secondary LCD and the “LOG” tag will
blink for a few seconds (see example below: record No. 27 dated July 14):
If there are less than 5 memory locations remaining, the record number
and the “Lo” message will be displayed, to alert the user.
If the log space is full, the “FULL LOG” message will be displayed and
no more data will be saved.
Along with the current measurement, the date, time, mV value,
temperature and calibration data is stored. If a HANNA P Type ORP
electrode is used pH information is not stored.
VIEW LOGGED DATAVIEW LOGGED DATA
VIEW LOGGED DATAVIEW LOGGED DATA
VIEW LOGGED DATA
Press RCL
while in measurement mode to retrieve the stored information.
If no data was logged, the instrument displays:
LOGGINGLOGGING
LOGGINGLOGGING
LOGGING
Notes: Press GLP to return to measurement mode.
If calibration has not been performed, the instruments
display “no CAL” message blinking.