Owner manual

30
NSL Requests the number of logged samples (4 chars).
LODxxx Requests the xxx
th
record logged data.
LODALL Requests all Log on demand data.
The answer string contains:
pH resolution (2 chars): 00 - pH 0.001 resolution, 01 - pH 0.01
resolution
• pH reading status (1 char): R - in range, O - over range, U -
under range
pH reading with sign and decimal point (7 chars)
Temperature reading sign and decimal point in ºC (7 chars)
mV reading status (1 char - R, O, U)
mV reading with sing and decimal point (7 chars)
the logged time, yymmddhhmmss (12 chars)
the calibration offset with sign and decimal point (7 chars)
the avarage of slopes with sign and decim al point (7 chars)
temperature probe presence (1 char)
Errors: “Err3” is sent if the Log on demand is empty.
“Err4” is sent if the requested set parameter is not available.
“Err5” is sent if an argument of the command is not correct.
“Err6” is sent if the requested range is not available.
“Err7” is sent if the instrument is in logging mode.
“Err8” is sent if the instrument is not in measurement mode.
Invalid commands will be ignored.