Owner's manual

Instruction Guide
Fixed-Volume Pipette
HI 731340
HI 731341
HI 731342
http://www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna Instruments product.
This instruction guide will provide you with the necessary
information for the correct use of the pipette. Please read it
carefully before using the pipette. If you need additional technical
information, do not hesitate to e-mail us at tech@hannainst.com.
Please examine this product carefully. Make sure that no
damage has occurred during shipment. If there is any
damage, notify your Dealer.
Each pipette is supplied complete with:
• Two tips
• Instruction Guide
• Rigid carrying case
Note: Keep all packing material until you did verify that the
pipette works correctly. Any defective item must be
returned in its original packing.
PRELIMINARY EXAMINATION
1. Use the pipette in a vertical position whenever possible.
2. Operate slowly and smoothly. Aspiration and dispensing should
been carried-out with the same speed.
3. The tip should be immersed in the sample liquid the minimum
necessary (2-4 mm). Greater depths may cause inaccurate
measures.
4. Change the tip whenever a different kind of sample is to be
aspired or when droplets remain on the inside wall of the tip
from the previous pipetting operation.
5. Work always with samples at room temperature.
1. Liquid should never enter the pipette shaft. To avoid this:
never use pipette without a tip
do not invert pipette upside down or lay it flat when there
is a liquid in the tip
press and release the push button slowly and smoothly.
2. When pipetting concentrated acids or alkalis or corrosive
solutions, pay particular attention to avoid contact of the
liquid with the pipette shaft, or it could be seriously damaged.
Extensive contact with corrosive fumes may result in premature
ageing of internal components of the pipette.
3. After use, store the pipette in its case or in a clean safe
place.
PIPETTE MAINTENANCE
GENERAL DESCRIPTION
HI 731340/1/2 is an air-displacement volumetric pipette
designed to measure and transfer fixed volumes of liquids
that needs to be precisely dosed.
The liquid is aspirated into deposable tips. The use of
disposable tips eliminate cross contamination of the liquid
samples.
To discard used tips, the pipette is equipped with a built-in
tip ejector to protect the user from contaminated tips.
HINTS FOR ACCURATE MEASUREMENTS
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.
ISTMICPIPR1 05/03
Hanna Instruments Srl
V.le delle Industrie 12/A
35010 Villafranca (Padova)
ITALY
http://www.hannainst.com
HI 731350 tips for 200 µL automatic pipette (25 pcs)
HI 731351 tips for 1000 µL automatic pipette (25 pcs)
HI 731352 tips for 2000 µL automatic pipette (4 pcs)
ACCESSORIES
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.