Instruction Manual HI 8014 • HI 8915 HI 931000 Portable pH Meters www.hannainst.
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the meter. This manual will provide you with the necessary information for a correct use of the instrument. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with the directives.
FUNCTIONAL DESCRIPTION HI8014 SPECIFICATIONS HI8014 1 2 Range pH pH meter HI 8014 °C pH OFF mV 3 °C mV/pH AP 4 ins rum n s 1) BNC socket for combination pH electrodes. 2) Liquid Crystal Display (LCD). 3) Rotary Knob with the following selectable positions: OFF To switch the meter off (in any other position the instrument is on). °C To display the manual temperature setting (through °C trimmer). pH To display the pH value.
FUNCTIONAL DESCRIPTION HI8915 7 1 2 3 pH HI 8915 ATC pHmeter ON OFF 8 HOLD pH 9 mV °C 4 OFFSET 5 6 1) 2) 3) 4) 5) 6) 7) 8) 9) 10) 11) 12) 6.86 7.00 + - mV OUTPUT 10 SLOPE 4.01 10.01 11 12 SPECIFICATIONS HI8915 Range pH mV °C Resolution pH mV °C Accuracy pH (@20°C/68°F) mV °C Typical EMC pH Deviation mV °C pH Calibration CALIBRATION BNC socket for combination pH or ORP electrodes. Liquid Crystal Dysplay (LCD). HOLD key, to freeze the reading on the display.
FUNCTIONAL DESCRIPTION HI931000 1 SPECIFICATIONS HI931000 Range ins rum n s Resolution 2 CAL ON OFF 3 °C 4 5 6 CFM Accuracy (@20°C/68°F) Typical EMC Deviation pH Calibration HI 931000 Temperature Compensation pH Electrode Battery Type Life Environment Dimensions 1) 2) 3) 4) 5) 6) DIN socket for combination pH/°C electrode. Liquid Crystal Display (LCD). ON/OFF key, to switch the instrument on or off. °C key, to display the measured temperature (press & hold).
pH MEASUREMENTS OPERATIONAL GUIDE INITIAL PREPARATION TOP VIEW °C BNC electrode connector OFF mV TOP VIEW OF HI 8915 pH Turn the rotary knob to the pH position to display the pH measurement. Shake the electrode briefly and allow a couple of minutes for the electrode to adjust and stabilize. The display will show the pH value compensated for the manually adjusted temperature. Turn the instrument ON, shake the electrode briefly and wait until the stability indicator "∆" stops blinking.
Notes: • In order to take accurate pH measurements, make sure that the instrument is calibrated for pH before use. • For a faster response time when using refillable electrodes, unscrew the refill hole cap. Dip the temperature probe or the pH/°C electrode into the sample and allow the reading to stabilize (a couple of minutes). For HI 931000, to display the measured temperature, press and hold the °C key. When the °C key is released, the display will return to the pH reading.
• Set the rotary knob to °C position to display the manual temperature setting and adjust the °C trimmer until the LCD shows the measured temperature. pH CALIBRATION For best accuracy, frequent calibration of the instrument is recommended. The instrument should be recalibrated for pH: a) Whenever the pH electrode or temperature probe is replaced. b) At least once a month. c) After testing aggressive chemicals. d) Whenever the battery has been replaced. e) When high accuracy is required.
• Stir gently and wait a couple of minutes for thermal equilibrium. Note: The electrode should be submerged approximately 4 cm (1½") into the solution. The temperature probe should be located as close as possible to the pH electrode. • Press °C to display the temperature of the buffer (e.g. 20.0°C). PROCEDURE FOR HI 931000 • Turn the meter on. Note: The "∆" indicator blinks when the reading is not stable. • Press CAL and the display will flash "7.01 ∆".
pH VALUES AT VARIOUS TEMPERATURES TEMPERATURE CALIBRATION For manual temperature compensation during calibration, please refer to the following chart. TEMP °C °F 32 0 41 5 50 10 59 15 68 20 77 25 86 30 95 35 40 104 45 113 50 122 55 131 60 140 65 149 70 158 4.01 4.01 4.00 4.00 4.00 4.00 4.01 4.02 4.03 4.04 4.05 4.06 4.07 4.09 4.11 4.12 6.86 6.98 6.95 6.92 6.90 6.88 6.86 6.85 6.84 6.84 6.83 6.83 6.84 6.84 6.85 6.85 pH VALUES 7.01 9.18 7.13 9.46 9.39 7.10 9.33 7.07 9.27 7.04 9.22 7.03 9.18 7.01 9.14 7.
ELECTRODE CONDITIONING AND MAINTENANCE If the bulb and/or junction are dry, soak the electrode in HI 70300 Storage Solution for at least one hour. For refillable electrodes: If the refill solution (electrolyte) is more than 2½ cm (1") below the fill hole, add HI 7082 (3.5M KCl Electrolyte Solution) for double junction or HI 7071 (3.5M KCl+AgCl Electrolyte Solution) for single junction electrodes. TEST MEASUREMENT Rinse the electrode tip with distilled water.
Inorganic Soak in Hanna HI 7074 Inorganic Cleaning Solution for 15 minutes. Oil /grease Rinse with Hanna HI 7077 Oil and Fat Cleaning Solution. IMPORTANT: After performing any of the above cleaning procedures rinse the electrode thoroughly with distilled water, drain and refill the reference chamber with fresh electrolyte (not necessary for gel-filled electrodes), and soak the electrode in HI70300 Storage Solution for at least 1 hour before taking any measurement.
ACCESSORIES pH CALIBRATION SOLUTIONS HI 70004P HI 7004M HI 7004L HI 7006M HI 7006L HI 70007P HI 7007M HI 7007L HI 7009M HI 7009L HI 70010P HI 7010M HI 7010L pH 4.01 buffer solution, 20 mL sachet, 25 pcs pH 4.01 buffer solution, 230 mL bottle pH 4.01 buffer solution, 500 mL bottle pH 6.86 buffer solution, 230 mL bottle pH 6.86 buffer solution, 500 mL bottle pH 7.01 buffer solution, 20 mL sachet, 25 pcs pH 7.01 buffer solution, 230 mL bottle pH 7.01 buffer solution, 500 mL bottle pH 9.
FC210B FC220B FC911B HI1413B Glass body, double junction, conic, Viscolene, nonrefillable combination pH electrode. Use: milk, yogurt. Glass body, triple ceramic, single junction, refillable, combination pH electrode. Use: food processing. PVDF body, double junction, refillable with builtin amplifier, combination pH electrode. Use: very high humidity. Glass body, single junction, flat tip, Viscolene, non-refillable combination pH electrode. Use: surface.
SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel. (21) 4584.2941 • Fax (21) 4584.2942 Japan: Tel. (03) 3258.9565 • Fax (03) 3258.9567 Korea: Tel. (02) 2278.5147 • Fax (02) 2264.1729 Malaysia: Tel. (603) 5638.9940 • Fax (603) 5638.9829 Singapore: Tel. 6296.