Instruction Manual HI 8427 - HI 931001 pH/mV Simulators pH/m HI 8427 V CA LIBR ATO R pH2 LOW BATT ERY LEAK AGE pH4 pH7 pH0 OFF pH12 pH14 1900 mV 0 pH10 -350 mV 350 mV 1900 mV 10 20 50 40 °C 30 HIGH IMPE DA NCE These Instruments are in Compliance with the CE Directives 1
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a more precise idea of its versatility in a wide range of applications. These instruments are in compliance with the directives EN 50081-1 and EN 50082-1.
FUNCTIONAL DESCRIPTION HI 8427 FUNCTIONAL DESCRIPTION HI 931001 9 HI 8427 pH/mV CALIBRATOR 7 LOW BATTERY LEAKAGE 8 pH2 pH4 pH7 pH10 3 3 1 2 pH0 pH12 OFF pH14 1900 mV -350 mV 350 mV 1900 mV 4 10 0 5 6 20 50 30 HIGH IMPEDANCE 40 °C 1. 2. 3. 4. 5. 6. 7. 8. 9.
working temperature. If not, adjust the pH meter to read the correct value. Note: The °C knob is not effective during pH 7 calibration, since pH 7 is used as reference. pH CALIBRATION HI 8427: • Connect the simulator to the pH meter using the HI 7858/1 cable provided. HI 8427 pH/mV CALIBRATOR pH • • • • • Switch the pH meter on and select pH 7 on the simulator. Make sure that the LOW BATTERY LED is not lit. If it is lit, change the battery of the simulator before performing the calibration.
• mV (ORP) CALIBRATION HI 8427: • Connect the simulator to the mV (ORP) meter using the HI 7858/1 cable provided. • HI 8427 pH/mV CALIBRATOR pH • Switch the mV meter on and select the desired mV value on the simulator. Make sure that the LOW BATTERY LED is not lit. If it is lit, change the bat-350 tery of the simulator bemV fore performing the calibration.
pH VALUES AT VARIOUS TEMPERATURE CABLE HIGH IMPEDANCE TEST HI 8427 only: • Connect the simulator to the cable (e.g.HI 7855). • Press the HIGH IMPEDANCE key and observe the LEAKAGE LED while you are turning the simulator rotary knob to any pH/mV value. • • If the LED does not light, the cable and connector going to the meter are in perfect condition. If the LED is lit, it means there is a leakage problem with the cable and connector and you should check/replace it.
ACCESSORIES BATTERY REPLACEMENT The instruments use a 9 volt battery that lasts for approximately 100 hours or 500 hours of continuous use with HI 8427 and HI 931001 respectively. In the event of dead battery the LOW BATTERY LED of HI 8427 will be lit or two additional decimal points will be displayed on HI 931001. HI 710001 Soft carrying case for HI 931001, dimensions 230 x 100 x 50 mm HI 7858/1 Extension cable with 3 mm diameter (0.12"), 1m (3.
WARRANTY CE DECLARATION OF CONFORMITY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrodes and the probes are warranted for a period of six months. This warranty is limited to repair or replacement free of charge. Damages due to accidents, misuse, tampering or lack of prescribed maintenance are not covered.