Instruction Manual HI 9064 - HI 9065 HI 9160 Portable Water-Resistant Hygrometers s H HI 9065 Thermo Hygrometer RH% 5-95% C 0-60 C F 32-140 F ON/OFF RH HOLD °C °F http://www.hannainst.
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with 1 and 50082-1.
FUNCTIONAL DESCRIPTION H I 9064 & H HII 9065 FUNCTIONAL DESCRIPTION H HII 9160 7 1 3 S H °C 8 9 1 e ins rum n s 4 S H °C S H 5 HI 9064 Hygrometer HI 9065 Thermo Hygrometer 6 RH% 5 . . . . . .
SPECIFICATIONS OPERATIONAL HI9064 & HI9160 HI9065 Range RH °C °F 5.0 to 95.0% RH --- 5.0 to 95.0% RH 0.0 to + 60.0 °C 32 to +140 °F Resolution RH °C °F 0.1% RH --- 0.1% RH 0.1 °C 1 °F Accuracy RH (@25°C/77°F) °C °F ±3% RH Typical EMC Deviation Calibration RH ±3% RH ±2 °C / ± 4°F Through 2 trimmers on RH probe HI70608/2 Probe Operating Conditions Dimensions Instrument RH Probe HI70605/2 INSTRUMENT PREPARATION Each meter is supplied complete with 4 each 1.5V AA batteries.
TEMPERATURE MEASUREMENTS (FOR HI9065 ONLY) Connect the HI 765N temperature probe into the socket placed on the top of the meter (refer to #9 in the Functional Description, page 4) and the unit will automatically display the readings from the temperature probe. For special applications, you can choose in a wide range of separate temperature probes (see Accessories at page 13). Note: HI 9065 has to be used with just one probe (RH or temperature) at a time.
2 H • Seal the chamber well when not in use, as the LiCl solution is extremely hygroscopic and tends to capture the humidity present in the air causing the solution to expand in volume and overflow from the container. Note: the central trimmer is related to the temperature calibration and you will find it on HI 9065 probe only. • Remove the probe and tightly seal the chamber containing the LiCl solution. O 12 cc • Pour approximately 12 cc of distilled water into the other chamber marked "RH 75.4%".
"BEPS" & BATTERY REPLACEMENT HI9064, HI9065 and HI9160 are supplied with the advanced "BEPS" Battery Error Preventive System technology that detects a low power condition in the battery and prevents erroneous measurements being taken due to low voltage. For HI 9064 and HI 9065, upon depletion of the power in the batteries up to a prefixed threshold, the LCD automatically switches off. For HI9160, when the batteries are low "V" is displayed on the LCD to warn the user.
WARRANTY All Hanna Instruments meters are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The probes are guaranteed for a period of six months. This warranty is limited to repair or replacement free of charge. Damage due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument.
HANNA LITERATURE WATER ANALYSIS HANDBOOK ENVIROCARE GENERAL CATALOG h t t p : / / w w w . h a n n a i n s t . c o m MANRHWPR2 06/96 These and many others catalogs, handbooks and leaflets are available from Hanna. To receive your free copy, contact your dealer or the nearest Hanna Customer Service Center.