ON
OFF
ZERO
MEASUREMENT PROCEDURE
Turn the meter on by pressing ON/OFF.
When the LCD displays "- - -", it is
ready.
Fill the cuvet with sample up to
the 10 mL mark and replace
the cap.
Place the cuvet into the holder
and ensure that the notch on
the cap is positioned securely
into the groove.
Press ZERO and "SIP" will appear on the display.
Wait for a few seconds and the
display will show "-0.0-". Now the
meter is zeroed and ready for mea-
surement.
Remove the cuvet and add
the content of one packet of
HI 93721 reagent. Replace
the cap and agitate until
dissolution is complete.
Reinsert the cuvet into the
instrument.
Press READ TIMED and the display will show the count-
down prior to the measurement or, alternatively, wait
for 3 minutes and press READ DIRECT. In both cases
"SIP" will appear during measurement.
The instrument directly displays concentration in mg/L of
iron on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Molybdate Molybdenum above 50 ppm
Calcium above 10,000 ppm (as CaCO
3
)
Magnesium above 100,000 ppm (as CaCO
3
)
Chloride above 185,000 ppm.
READ
TIMED
12
3
6
9
3’
READ
DIRECT
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
Do not touch the cuvet walls with hands.
In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
It is important that the sample does not contain any
debris. This would corrupt the readings.
It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
All the reaction times reported in this manual are
referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved at
30°C (86°F).
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLID
E
OFF
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environ-
ment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences
to radio and TV equipments, requiring the operator to take all necessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
REAGENT SETS
HI 93721-01 Reagents for 100 tests
HI 93721-03 Reagents for 300 tests
OTHER ACCESSORIES
HI 710009 Blue rubber boot
HI 710010 Orange rubber boot
HI 721310 9V battery (10 pcs)
HI 731318 Tissue for wiping cuvets (4 pcs)
HI 731321 Glass cuvets (4 pcs)
HI 731325 Caps for cuvets (4 pcs)
HI 93703-50 Cuvets cleaning solution (230 mL).