User Guide
9
the Standard 7.01 or the NIST 6.86 are the most appropriate for this
purpose even though both meters can be calibrated with any of the 5
memorized calibration solutions.
• Immerse the pH electrode into a pH buffer
solution (e.g. pH 6.86) and stir gently.
Note: the electrode should be submerged ap-
proximately 4 cm (1½") into the
solution. With the temperature probe
located as close to it as possible.
• Press the CAL key. The "CAL" and "
1
BUF
" indicators will be dis-
played. The secondary LCD will display one of the 5 buffers
(e.g."4.01").
• Press the
Ï
°C or Ð
°C°C
°C°C
°C key until the secondary display changes
to the buffer (“6.86”).
• Once the electrode is submerged in the buffer solution, the LCD
will notify the user if the reading is not stable by an intermittent
"NOT READY". Only when the reading is stable, will it change to a
blinking "READY" and "CON".
• Press the CFM key to confirm the calibration. If the reading is not
close to the selected buffer, "WRONG
1
BUF
" and "WRONG " will
blink alternatively.
If the reading is close to the selected buffer, the meter stores the
reading (and adjusts the offset point). The buffer value is then
displayed on the primary LCD and the
secondary LCD will display another buffer
value (e.g."4.01").
HI 7006
CAL
1
BUF
pH
CAL
CAL
1
BUF
pH
CAL
1
BUF
pH
pH
WRONG
WRONG
2
BUF
CFM
NOT READY
READY
CON










