mometer), e.g. 20°C.
Set the rotary knob to °C to display the
manual temperature setting.
Adjust the °C trimmer until the LCD shows the measured tem-
perature.
Set the rotary knob to pH to display pH
measurement and wait for the reading
to stabilize.
Adjust the AP trimmer with a small screwdriver until the LCD
shows the pH value at the above temperature (see the pH versus
temperature chart on page 35).
Rinse and immerse the pH electrode in
pH 4.01 or pH 10.01 buffer (2
nd
calibra-
tion point) and stir gently.
Wait a couple of minutes and adjust the mV/pH trimmer until
the LCD shows the pH value at the noted temperature (see
page 35 for pH versus temperature chart).
The pH calibration is now complete.
PROCEDURE FOR HI 8314 & HI 8915
Switch the meter on after connecting the
pH electrode and the temperature probe.
OFF
mV
°C
pH
°C
ºC
OFF
mV
°C
pH
A
P
pH
°
C
HI 7004
pH
mV/pH
ON
OFF
31
Rinse the pH electrode with some pH 4.01
or 10.01 buffer solution (2
nd
calibration
point).
Dip the electrode into the pH 4.01 (or
pH 10.01) buffer solution.
Stir gently and wait until the stability
indicator "" stops flashing.
Press CFM to confirm the acceptance of
the second buffer solution.
pH calibration is now complete and the meter
returns to the measurement mode.
PROCEDURE FOR HI 931000
Turn the meter on.
Note: The stability indicator "" blinks when
the electrode reading is unstable.
Press CAL and the display will flash
"7.01 ".
Remove the protective cap from the elec-
trode, rinse the tip of the electrode with
some pH 7.01 solution, then immerse it
in a pH 7.01 buffer solution. Stir gently
and wait a couple of minutes for thermal
equilibrium.
Note: The electrode should be submerged
approximately 4 cm (1½") in the
solution.
When electrode has stabilized, the sta-
bility indicator "" stops flashing.
Press CFM to confirm the first buffer solution.
in
s
e
r
u
m
n
s
O
N
O
F
F
C
AL
C
F
M
HI 9214 stick pHmeter
HI 7004
ON
OFF
CAL
CFM
HI 7007
4 cm
(1½
")
CFM