Instruction Manual HI 9828 Multiparameter w w w . h a n n a i n s t .
Dear Customer Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instrument. It will provide you with the necessary information for correct use of the instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com or see the back cover for our worldwide contact list. This instrument is in compliance with the directives.
TABLE OF CONTENTS CHAPTER 1- INTRODUCTION 1.1 Preliminary examination .............................................................................. 7 1.2 Model identification .................................................................................... 7 1.3 General description .................................................................................... 8 1.4 Power supply .............................................................................................. 9 1.4.
4.3 Dissolved oxygen calibration ..................................................................... 33 4.3.1 Procedure .......................................................................................... 33 4.4 Conductivity calibration ............................................................................ 34 4.4.1 Procedure .......................................................................................... 35 4.5 Atmospheric pressure ....................................................
APPENDIX A - PROBE MAINTENANCE ............................................................................ 53 B - ACCESSORIES .......................................................................................... 55 C - WARRANTY .............................................................................................. 57 USER NOTES ............................................................................................
Chapter 1 - INTRODUCTION 1.1 PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer or the nearest Hanna Customer Service Center immediately. HI9828 is supplied complete with: - Multisensor probe (pH/ORP, Conductivity, D.O.
1.3 GENERAL DESCRIPTION HI 9828 is a multiparameter system that benefits from years of experience of Hanna Instruments as a manufacturer of analytical instruments. Waterproof, resistant and easy to use, it is the ideal solution for field measurements of lakes, rivers and sea.
1.4 POWER SUPPLY HI9828 runs with 4 rechargeable C size, Ni-MH batteries. On the display the battery symbol visualizes the remaining battery charge. When this symbol starts blinking, it is necessary to recharge or replace them with new ones. When the batteries are completely rundown the meter automatically shuts off to avoid wrong readings. 1.4.1 Batteries installation Battery replacement must only take place in a nonhazardous area.
Cigarette lighter supply To recharge HI9828 from a vehicle cigarette lighter supply, use HI710046 HI710046. • Simply connect the cable to the meter and to the car’s cigarette lighter. • The message “Battery charging in progress” appears on the display, followed by the battery symbol. • A complete batteries recharging will last about 14 hours. NO TE NOTE Batteries can also be recharged with the meter ON; if the auto-off feature is enabled, the meter turns off automatically after the set time. 1.
Since the Pt/PtO system depends on the pH, the pretreatment of the electrode may be determined by the pH and the redox potential values of the solution to be measured. As a general rule, if the ORP (mV) reading corresponding to the solution pH value is higher than the values in the table below, an oxidizing pretreatment is necessary; otherwise perform a reducing pretreatment.
1.5.3 D.O. sensor activation The D.O. probe is shipped dry. To hydrate the probe and prepare it for use proceed as follows: • Remove the black & red plastic cap. This cap is used for shipping purposes only and can be thrown away. • Insert the supplied O-ring in the membrane. • Rinse the supplied membrane with electrolyte while shaking it gently. Refill with clean electrolyte. Gently tap the membrane over a surface to ensure that no air bubbles remain trapped.
1.6 SPECIFICATIONS TEMPERATURE Range Resolution Accuracy Calibration pH Range Resolution Accuracy Calibration ORP Range Resolution Accuracy Calibration DISSOLVED OXYGEN Range Resolution Accuracy Calibration -5.00 to 55.00 °C; 23.00 to 131.00 °F; 268.15 to 328.15 K 0.01 °C; 0.01 °F; 0.01 K ± 0.15 °C; ± 0.27 °F; ±0.15 K Automatic at 1 custom point 0.00 to 14.00 pH; ± 600.0 mV 0.01 pH; 0.1 mV ± 0.02 pH; ± 0.5 mV Automatic 1, 2 or 3 points with 5 memorized standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.
CONDUCTIVITY Range 0.000 to 200.000 mS/cm (actual EC up to 400 mS/cm) Resolution Manual Automatic 1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm 1 µS/cm from 0 to 9999 µS/cm 0.01 mS/cm from 10.00 to 99.99 mS/cm 0.1 mS/cm from 100.0 to 400.0 mS/cm Automatic mS/cm 0.001 mS/cm from 0.000 to 9.999 mS/cm 0.01 mS/cm from 10.00 to 99.99 mS/cm 0.1 mS/cm from 100.0 to 400.
SALINITY Range Resolution Accuracy Calibration 0.00 to 70.00 PSU (extended Practical Salinity Scale) 0.01 PSU ±2% of reading or ±0.01 PSU whichever is greater 1 custom point SEAWATER SPECIFIC GRAVITY Range 0.0 to 50.0 σt, σ0, σ15 Resolution 0.1 σt, σ0, σ15 Accuracy ± 1σt, σ0, σ15 Calibration Based on conductivity or salinity calibration ATMOSPHERIC PRESSURE Range Resolution Accuracy Calibration 450 to 850 mmHg; 17.72 to 33.46 inHg; 600.0 to 1133.2 mbar; 8.702 to 16.436 psi; 0.5921 to 1.1184 atm; 60.
GENERAL CHARACTERISTICS Temperature Compensation automatic from -5 to 55 °C (23 to 131 °F) Logging Memory Up to 60,000 samples with 13 measurements each* Logging Interval From 1 second to 3 hours PC Interface USB (with HI 92000 software) Waterproof PProtection rotection Meter IP67, Probe IP68 Environment 0 to 50 °C (32 to 122 °F); RH 100 % Power Supply 4 x 1.5 V alkaline C cells (approximately 150 hours of continuous use, without backlight) or 4 x 1.
1.7 DISPLAY & KEYBOARD DESCRIPTION 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12.
1.8 HELP FUNCTION HI9828 is provided with the HELP function, useful to have short information regarding the displayed screen. Simply press the HELP key and an information window will appear. For longer messages press the arrow keys to scroll. To escape from the help window press again the HELP key or ESC.
Chapter 2 - MEASUREMENT MODE HI9828 can read at the same time different parameters from the same probe. As described in the previous section, up to 3 sensors can be mounted on the probe. 2.1 PROCEDURE • Connect the probe to the meter and carefully fix the protection sleeve to the probe. • Immerse the probe into the sample while paying attention to avoid stones. • Turn the meter on by pressing the On/Off key. The display will show “HANNA HI9828”, the firmware version and then enters the measurement mode.
Chapter 3 - SETUP MODE A few parameters have to be set before taking any measurements. In the main menu two setup items are available: “Measurement” and “System”. The measurement setup allows to set the displayed readings and their units, while the system setup is used to set the system parameters, as the interface language, date and time, LCD contrast, acoustical signals, etc. 3.1 MEASUREMENT SETUP • Switch the meter on by pressing On/Off.
pH, mV of pH Input, ORP, D.O. % Saturation, Salinity These parameters can only be enabled or disabled; the measure unit and the resolution are fixed. D.O. concentration It is possible to select ppm or mg/L measure unit. Conductivity and Actual Conductivity It is possible to select among the following options: Auto (autoranging both µS/cm and mS/ cm ranges), 1 µS/cm, 0.001 mS/cm, 0.01 mS/ cm, 0.1 mS/cm, 1 mS/cm, Auto mS (autoranging for mS/cm ranges).
NO TE NOTE A maximum of 12 measurements can be displayed simultaneously. A warning message appears if trying to enable more than 12 measurements. 3.2 SYSTEM SETUP • From measurement mode, press MENU, select “System Setup” using the arrow keys and then press OK. • To select a parameter, scroll with the arrow keys to highlight it and then press MODIFY. NO TE NOTE If the password is enabled, the meter will ask to insert it prior to change the first parameter.
Use the keyboard to insert the time and then press OK. To choose AM or PM, press A or P on the keyboard after inserting the time. Auto off The meter shuts off automatically if no keys are pressed for the set time. By pressing the MODIFY softkey it is possible to select the desired option for this feature: NO (disabled), 5, 10, 15, 20, 30 or 60 minutes. Log interval Set the logging time interval from 1 second to 3 hours.
TDS unit TDS readings can be displayed in ppm-ppt or mg/L-g/L unit. Press MODIFY to select the desired option. Average length In order to obtain an average and more representative measurement with unstable samples, set a reading repetition number for the displayed parameters. To select the desired average length, press MODIFY. The value can be set from 1 to 30. Key beep If enabled, the meter emits an acoustical signal every time a key is pressed.
• The meter returns to the “System Setup” menu and the checkbox near the password entry is signed. To disable the password proceed as follows: • Press MODIFY to select the password setup item. • Insert the password and then press DISABLE. “NO” will appear in the text box. • Press OK to confirm. Restore factory settings It is possible to reset the “System Setup” and “Measurement Setup” parameters to their default values. • Select the “Restore factory settings” item and press OK.
3.3 TABLE OF MEASUREMENT AND SETUP ITEMS Measurement Setup Item Description Default value Valid Values Temperature Temperature unit °C K; °C; °F; pH pH measure 5 5; mV of pH input mV of pH readings 5 5; ORP Redox measure 5 5; D.O. % saturation Dissolved oxygen measure 5 5; D.O. concentration Dissolved oxygen measure ppm ppm; mg/L; Conductivity Electrical conductivity measure Auto ; Auto; 1 µS; 0.001 mS; 0.01 mS; 0.
Ref. temperature Reference temperature for conductivity measurements 25°C 20°C; 25°C Temp. coefficient Temperature coefficient for conductivity measurements 1.90%/°C 0.00 to 6.00%/°C 0.50 0.00 to 1.
Chapter 4 - CALIBRATION MODE HI9828 allows to perform six different types of calibration, one for each parameter and also a quick single-point calibration for some parameters. The calibration data are stored in the non volatile probe memory, so that the same probe can be used with different meters without needing recalibration. • To perform a calibration, select “Calibration” in the main menu with the arrow keys and then press OK.
• When the measure is stable, “Ready” is shown. Press CONFIRM to store the value. • The messages “Storing data on probe, please wait...” and “Updating GLP data, please wait ...” appear. NO TE If pH calibration is not required, the meter allows to skip to the EC quick NOTE calibration, by pressing the SKIP soft key. If the pH sensor is missing the message “pH sensor not installed! Skip to conductivity calibration” appears. • After the pH calibration is completed, the “Conductivity” option will blink.
4.2 pH CALIBRATION Calibrate the meter often, especially if high accuracy is required. Selecting pH calibration the display shows two options: “Calibrate pH” and “Clear old calibration”. If “Calibrate pH” is selected, it is possible to perform a new calibration at 1, 2 or 3 points with standard buffers (pH 4.01, 6.86 or 7.01, 9.18 or 10.01) or a single calibration with custom buffer. If “Clear old calibration” is selected, all calibration data will be deleted and the default data restored.
• Press CONFIRM to accept the value or BUFFER to select another buffer using the arrow keys. • After the first calibration point is confirmed, immerse the probe in the second buffer solution and stir gently. • When the reading is stable and close to the selected buffer, the display shows the message “Ready”. • Press CONFIRM to accept the value or BUFFER to change the buffer.
4.2.3 Error list If the meter does not accept a pH calibration point, a short message is displayed to indicate the possible error source. See for example the following screens: These are the available messages: • “Input out of scale”: the pH value is out of scale. • “Wrong buffer”: this message means that the difference between the pH reading and the selected buffer is too big. Check if the proper calibration buffer has been selected.
4.3 DISSOLVED OXYGEN CALIBRATION If D.O.% saturation range is calibrated, D.O. concentration range will also be calibrated, and vice-versa. The D.O.% saturation value is referred to the D.O. concentration in air (100%). For this reason it is recommended to calibrate the probe near the area where the measurements will be taken. Also note that the D.O. concentration values are based on D.O.% saturation, temperature, salinity and atmospheric pressure.
• Press CONFIRM to store the calibration point. • The following messages will appear: “Storing data on probe, please wait...”, “Updating GLP data, please wait ...” and “Calibration completed”. • Press OK to return to the calibration menu. • To return to the main menu, press ESC repeatedly. NO TE NOTE The user can perform a single point calibration with standard values. To abort the calibration press ESC after the first point is accepted. If the D.O.
The 3 options are related, so that one of this calibration procedure also calibrate the two remaining ranges. NO TE NOTE For correct EC readings it is recommnded to calibrate using a standard solution with a conductivity value close to the sample being measured. 4.4.1 Procedure After choosing the conductivity calibration mode in the main calibration menu, select the type of calibration with the arrow keys and then press OK. Conductivity • Select the “Conductivity” option and press OK to confirm.
Actual Conductivity • Select the “Actual conductivity” option and press OK to confirm. • Insert the custom value and set its resolution. • Immerse the probe in the conductivity solution and wait for stable reading. The probe sleeve must be inserted. • When the reading is stable, press CONFIRM to store the value. • After confirmation, the following messages are displayed: “Storing data on probe, please wait...”, “Updating GLP data, please wait...”, “Calibration complete”.
4.5 ATMOSPHERIC PRESSURE For this calibration procedure a reference barometer is needed. A maximum difference of 40 mbar between current reading and calibration point is allowed during calibration. 4.5.1 Procedure Choose the “Atmospheric pressure” calibration mode in the calibration menu, then select the calibration type using the arrow keys and confirm the selection by pressing OK. • To perform a pressure calibration at a custom point, select the “Custom pressure” option.
• When the reading is stable, press CONFIRM to store the calibration point. • After confirmation, the following messages are displayed: “Storing data on probe, please wait...”, “Updating GLP data, please wait...”, “Calibration complete”. • Press OK to return to the main calibration menu. • To return to the main menu, press ESC repeatedly. • To restore the factory calibration data, select the corresponding option in the “ORP calibration” menu and then press OK. 4.
Chapter 5 - LOGGING MODE HI9828 can store up to 60,000 samples in 100 different lots. The value 60,000 is reached if no remarks are used. When using remarks, the maximum number of stored measurements decreases, but in practical cases it will never be less than 50,000. 5.1 LOGGING • From measurement mode press LOG to store the enabled readings. The default setting suggests the last used lot to store the sample. Each sample can be associated to a tag by simply touching the tag with the tag reader.
ists, a waring message advises the user: “The file already exists! Insert a different file name”. Press OK to insert a different file name. • The “Add remark? Yes or No” window will then appear. If YES is pressed and a remark list already exists, it is possible to select the desired note or press NEW to insert a new remark in a text box appears. • The “Tag reading” option allows to associate the logged or to be logged samples to a tag. The message “Touch the tag with the tag reader” is displayed.
5.2.1 Lots This option allows to insert a new lot, to view logged measurements, to plot data or to delete lots. • Scroll with the arrow keys to select the desired lot and then press OK. • To create a new lot press the NEW LOT softkey and insert the identification name with the keyboard. Press OK to confirm. NO TE In the upper line of the window, the disNOTE play shows the percentage of memory still available for inserting new data, for example “Data lots (free: 100%)”.
Plot • Press PLOT and the list of the available parameters for the selected lot will appear. • Use the arrow keys to scroll and select the desired parameter. Press OK to view the graph. • Use the arrow keys to move the cursor in the graph and highlight a sample. The sample data are displayed below the graph. • Press ESC to return to the parameter list. • Press ESC again to return to the 3-option menu. NO TE NOTE The number of lot samples that can be plotted is limited by the display resolution.
5.2.4 Delete all remarks • To delete all the inserted remarks, use the arrow keys to select the entry and press OK. The display will shows the message “All stored remarks will be erased! Continue?”. Press YES to delete or NO to return to the previous screen. 5.2.5 Tags Read tag • Select READ TAG to view and modify the information associated to a tag, or to insert a tag ID if the tag has never been memorized in the meter. The display shows the message “Touch the tag with the tag reader”.
ID Æ S/N This option allows to view the tag serial number related to an ID. • Select IDÆS/N and press OK. • Insert the identification code using the keyboard and then press OK. • The tag information window will appear. Press OK to return to the previous screen or MODIFY to modify the tag ID. NO TE NOTE If the inserted ID is not present in the memory, a warning message advises the user. Add tag manually An ID code can be associated to a tag even if the tag is not physically available.
Chapter 6 - GLP GLP (Good Laboratory Practice) is a set of functions that allows to store or recall data regarding the probe calibration. This feature also allows to associate a reading to “certified data” (standard solutions, reference meters, etc.) put in the meter through the calibration procedure. To visualize GLP data, from measurement mode press the MENU softkey and scroll using the down arrow key to highlight the “GLP data” option. Press OK: the complete list of available parameters appears.
6.2 pH • From the GLP data main menu, select the pH option and press OK. • All the information about the last pH calibration appears: offset, acidic slope, basic slope, used buffers, time and date of procedure. • Use the arrow keys to scroll the last 5 calibration data sets stored. • Press ESC to return to the GLP data main menu. NO TE NOTE The C label near the buffer value indicates a custom point, while the H means HANNA standard value.
6.4 CONDUCTIVITY • From GLP data main menu select the “Conductivity” option and press OK. This menu allows to view data for conductivity, actual conductivity and salinity calibrations. • All the information about the last conductivity calibration appears: calibrated point, cell constant value, calibration type (conductivity, actual conductivity or salinity), time and date. • Use the arrow keys to scroll the last 5 memorized calibrations.
6.6 ORP • From GLP data main menu select the “ORP” option and press OK. • All the information about the last ORP calibration appears: calibrated point, time and date • In case of restoring factory calibration data, the display will show the warning message “Factory calibr. restored”. • Use the arrow keys to scroll the last 5 memorized calibrations. NO TE NOTE If no ORP calibration has been performed, a warning message advise the user. Press OK to return to the previous screen.
Chapter 7 - PC CONNECTION MODE The logged data can be transferred to PC using the HI92000 Windows® compatible application software. HI92000 allows to use the powerful capabilities of most spread sheet programs (e.g. Excel©, Lotus 1-2-3©). Simply open the file downloaded by HI92000 from the selected spread sheet program and you can do any elaboration available with the software (e.g. graphics, statistic analysis).
Chapter 8 - ERROR MESSAGES HI9828 displays a series of messages if probe or meter errors are generated. Here below are described the error messages with their meaning and indication to solve the problem. For quick information the help menu is always available on the meter by pressing the “Help” button. • “Continuous logging - Flash memory is full”. This message means that the memory is full and no reading can be logged. Press OK and delete one or more lots. • “Flash memory error!”.
• ”Probe critical error”. This message means that a probe generic error occurs. Turn the meter off, disconnect the probe and connect it again. If the problem persists, contact the HANNA service center. • “None of the enabled measurements is available”. This message appears in measurement mode if none of the selected parameters is available because the corresponding sensors are not mounted on the probe. Turn the meter off, install the required sensor and proceed with measurements.
APPENDIX A - PROBE MAINTENANCE HI9828 is supplied complete with a probe maintenance kit that includes HI7042S (electrolyte solution for D.O. sensor), 5 membranes and O-rings for D.O. sensor, a small brush for cleaning EC and D.O. sensors, 5 O-rings for sensor connectors and a syringe with grease to lubricate these O-rings. General maintenance After use rinse the probe with tap water and dry it. The pH electrode bulb must be kept moist. Dry the D.O. and EC sensors. Maintenance of D.O.
• Inspect the electrode for scratches or cracks. If any is present, replace the electrode. • Cleaning procedure: clean frequently the probe by soaking it for 1 minute in HI70670 or HI70671 cleaning solution. After cleaning soak the electrode in HI70300 storage solution before taking measurements. Maintenance of EC probe • After every series of measurements, rinse the probe with tap water. • If a more thorough cleaning is required, clean the probe with the supplied brush or a nonabrasive detergent.
APPENDIX B - ACCESSORIES HI 9828 ACCESSORIES HI 769828/4 HI 769828/10 HI 769828/20 HI 769828-0 HI 769828-1 HI 769828-2 HI 769828-3 HI 9828-25 HI 9828-27 HI 7698281 HI 92000 HI 920005 HI 7698282 HI 7698283 HI 7698284 HI 710045 HI 710046 HI 710005 HI 710006 HI 710012 HI 710013 Multiparameter probe body with 4 m cable Multiparameter probe body with 10 m cable Multiparameter probe body with 20 m cable pH sensor, double junction, non refillable pH/ORP sensor DO sensor EC sensor Quick calibration solution, 500 m
HI 5008 HI 5009 HI 5091 pH 8.00 buffer solution, 500 mL bottle pH 9.00 buffer solution, 500 mL bottle pH 9.18 buffer solution, 500 mL bottle HI 5010 pH 10.
APPENDIX C - WARRANTY All Hanna Instruments meters are guaranteed for two years (sensors, electrodes and probes for six months) against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. This warranty is limited to repair or replacement free of charge. Damage due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument.
USER NOTES 58
SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel. (21) 4584.2941 • Fax (21) 4584.2942 Japan: Tel. (03) 3258.9565 • Fax (03) 3258.9567 Korea: Tel. (02) 2278.5147 • Fax (02) 2264.1729 Malaysia: Tel. (603) 5638.9940 • Fax (603) 5638.9829 Singapore: Tel. 6296.7118 • Fax 6291.