User Guide
12
the last printed sample number. It is then possible to select
another sample.
• Press the ALT and FNC keys to return to setup mode.
Viewing logged data
• Press CH/TIME (HI98804 only) or TEMP/TIME (HI98801 only) to
view data of the selected sample. Data will be displayed in the
following order:
channel 1 temperature reading
channel 2 temperature reading (HI 98804 only)
channel 3 temperature reading (HI 98804 only)
channel 4 temperature reading (HI 98804 only)
date
time
• If CH/TIME or TEMP/TIME is pressed when the time is displayed,
the LCD will pass to the next Sample number.
• It is then possible to scroll the data of the next sample by pressing
CH/TIME or TEMP/TIME or select a different sample by the arrow
keys.
• to return to setup mode, press ALT and FNC when the meter
displays the sample number.
Cod. 02 - Lot summary printout
• Select the code 02.
• The message "Lot table Printing" will scroll twice across LCD.
• The meter will then print a complete set of information based on
the data stored in memory:
a - lot number
b - logging interval
c - starting date
d - number of samples
e - starting time
f - logged channels numbers.
Note For lot 00 the number of samples only will be printed.
DELETE LOGGED DATA (HI98801 and HI98804 only)
• Select code 05 to delete the Log on Demand data or code 06 to
delete the Timed Log Data.
• A scrolling message will be displayed.
• Press the ALT and CFM keys to confirm deletion.
• It is also possible to escape without data deletion pressing the ALT
and FNC keys.