PROBE MAINTENANCE
1. Unplug the HI764113 from the HI98198 meter. Remove the
stainless steel guard from probe body and set aside. Wipe off the
probe and cap with a soft tissue.
2. Remove the expired “Smart Cap” from the probe; Squeeze the
used cap at the notched V and pull it off the probe body (without
twisting).
3. Remove the used o-ring from the o-ring groove by rolling it off the
probe.
4. Wipe the o-ring groove with a soft tissue followed by a cleaning with
the supplied lens cleaner wipe.
5. Remove the new o-ring from the HI764113-1 kit and slide on probe
body using care not to roll or twist the o-ring.
6. Remove the syringe plunger; cut top off supplied sachet with silicone
grease and empty contents into the syringe. Using the syringe,
sparingly lubricate the o-ring with a thin film of the supplied grease.
Avoid getting any kind of grease or fingerprints onto the optical
window.
Do not substitute other grease/lubricants as it may cause the o-ring
to swell.
7. Clean the lens with an unused portion of the supplied lens cleaner wipe.
8. Remove the new optical cap from the box. Align the notched cutout
arrow on the Smart Cap with the matching guide on the probe body.
9. Slide and press the Smart Cap onto the HI764113 body until
the cap snaps in place. Once the cap is installed, it should not be
removed unless a new cap is required.
10. Connect the HI764113 probe to the HI98198 meter by
connecting the connector to the DIN jack located on the
top of the meter.
11. Power the meter to initiate the cap timer.
12. Calibrate.
13. Reinstall the stainless steel guard onto the probe body.