Instruction Manual HI 9835 Waterproof EC/TDS/NaCl/°C Meter for Lab and Field www.hannainst.
Dear Customer, Thank you for choosing a HANNA instruments® product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for correct use of the instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. This instrument is in compliance with the directives.
SPECIFICATIONS FUNCTIONAL DESCRIPTION Range EC 0.00 to 29.99 μS/cm (Autoranging) 30.0 to 299.9 μS/cm 300 to 2999 μS/cm 3.00 to 29.99 mS/cm 30.0 to 200.0 mS/cm up to 500.0 mS/cm actual(*) conductivity TDS 0.00 to 14.99 ppm (Autoranging) 15.0 to 149.9 ppm 150 to 1499 ppm 1.50 to 14.99 g/L 15.0 to 100.0 g/L up to 400.0 g/L actual(*) TDS (with 0.80 factor) NaCl 0.0 to 400.0 % Temp. 0.0 to 60.
EC Calibration NaCl Calibration Temperature Calibration Temperature Compensation Temperature Coefficient TDS Factor Probe Auto-off Power supply Casing Environment Dimensions Weight 1 point, with 6 memorized buffers 84,1413,5000,12880,80000,111800 μS/cm 1 point, with HI 7037 buffer (optional) 2 point, at 0 and 50°C (plus ±1°C adjustment) Automatic or Manual from 0 to 60°C (can be disabled to measure actual conductivity) Adjustable from 0.00 to 6.00 %/°C (EC and TDS only); default value is 1.
AUTORANGING EC/TDS CALIBRATION The EC and TDS scales are autoranging. The meter automatically sets the scale with the highest possible resolution. By pressing ALT+FIXED, the autoranging feature is disabled and the current range is frozen on the LCD. "F1" symbol blinks on the LCD. To restore the autoranging option press ALT+FIXED again. Note: Autoranging is automatically restored if the RANGE key is pressed, if the setup or calibration modes are entered and if the meter is turned off and back on again.
III IV V Use plastic beakers to minimize any EMC interference. The meter uses 1.90%/°C compensation factor during calibration. If the setup item "tc" has been set to a different value, when exiting the calibration mode the value displayed on the upper LCD could be different from the solution nominal value. It is possible to set the cell constant value directly without following the calibration procedure. Enter the setup mode and select "CEL" (see "Setup" section for details).
CONDUCTIVITY VS TEMPERATURE CHART SETUP The conductivity of an aqueous solution is the measure of its ability to carry an electrical current by means of ionic motion. The conductivity invariably increases with increasing temperature. It is affected by the type and number of ions in the solution and by the viscosity of the solution itself. Both parameters are temperature dependent.
The following table lists the setup items, their valid range of values and the factory settings (default): Item tc tcE tdS CEL Aof vEr Chr Description Temp. compensation coeff. Temp. compensation mode TDS factor Cell constant (K) Auto-off enable Firmware release Battery level test Valid values Default 0.00 to 6.00 %/°C 1.90 Atc, Mtc, notc Atc 0.40 to 0.80 0.50 0.500 to 1.700 1.000 On, OFF On Notes I Once enabled, the Auto-off time is fixed at 5 minutes.
ACCESSORIES CONDUCTIVITY CALIBRATION SOLUTIONS OTHER ACCESSORIES HI 70030P HI 7030L HI 7030M HI 70031P HI 7031L HI 7031M HI 70033P HI 7033L HI 7033M HI 7034L HI 7034M HI 7035L HI 7035M HI 70039P HI 7039L HI 7039M 12880 μS/cm solution, 20 mL sachet (25 pcs.) 12880 μS/cm solution, 500 mL bottle 12880 μS/cm solution, 230 mL bottle 1413 μS/cm solution, 20 mL sachet (25 pcs.) 1413 μS/cm solution, 500 mL bottle 1413 μS/cm solution, 230 mL bottle 84 μS/cm solution, 20 mL sachet (25 pcs.
WARRANTY CE DECLARATION OF CONFORMITY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrodes and the probes are warranted for a period of six months. This warranty is limited to repair or replacement free of charge. Damages due to accidents, misuse, tampering or lack of prescribed maintenance are not covered.
SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel. (21) 4584.2941 • Fax (21) 4584.2942 Japan: Tel. (03) 3258.9565 • Fax (03) 3258.9567 Korea: Tel. (02) 2278.5147 • Fax (02) 2264.1729 Malaysia: Tel. (603) 5638.9940 • Fax (603) 5638.9829 Singapore: Tel. 6296.