For HI8417 and HI8521
Connect the ORP electrode to the meter and submerge the tip (4
cm / 1½") into the sample to be tested.
Note: ORP measurements are taken without temperature compensa-
tion.
Press the pH/mV key (HI8417) or
RANGE key (HI8521) once to enter
the mV mode.
Allow a few minutes for the reading to
stabilize.
The display will indicate the absolute
mV value (positive or negative).
When values are less than ±400 mV, tenths of mV are dis-
played; for values greater than ±400 mV, the decimal digit
disappears. Note that the change in scale (from 0.1mV to 1mV
resolution) is automatic. "Err2" for HI8417 or "E3" for HI8521
appears if the value exceeds ±1999 mV (out of range).
AFTER MEASUREMENTS
Press the ON/OFF key again to switch the instru-
ment OFF.