PH VALUES AT VARIOUS TEMPERATURE
Temperature has an effect on pH. The calibration buffer solutions are
effected by temperature changes to a lesser degree than normal
solutions.
For manual temperature calibration (standard with HI 8519 or op-
tional with HI 8417, HI 8520 and HI 8521) please refer to the
following chart.
For instance, if the buffer temperature is 25°C, the display should
show pH 4.01, 7.01 or 10.01 at pH 4, 7 or 10 buffers, respectively.
At 20°C, the display should show pH 4.00, 7.03 or 10.06. The meter
reading at 50°C will then be 4.06, 6.98 or 9.82.
TEMP pH VALUES
°C °F 4.01 6.86 7.01 9.18 10.01
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
32
41
50
59
68
77
86
95
104
113
122
131
140
149
158
4.01
4.00
4.00
4.00
4.00
4.01
4.02
4.03
4.04
4.05
4.06
4.07
4.09
4.11
4.12
6.98
6.95
6.92
6.90
6.88
6.86
6.85
6.84
6.84
6.83
6.83
6.84
6.84
6.85
6.85
7.13
7.10
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
6.99
6.99
9.46
9.39
9.33
9.27
9.22
9.18
9.14
9.10
9.07
9.04
9.01
8.99
8.97
8.95
8.93
10.32
10.24
10.18
10.12
10.06
10.01
9.96
9.92
9.88
9.85
9.82
9.79
9.77
9.76
9.75
TEMPERATURE CALIBRATION
(HI8417, HI8520, HI8521 ONLY)
All Hanna bench-top meters have been accurately temperature cali-
brated at the factory. We recommend that you have your bench-top
meters temperature calibrated once a year.
Your nearest Hanna Service Center is available to quickly service your
meter. Contact your Dealer or the nearest Hanna Service Center for
more information.