HANNA LITERATURE Instruction Manual Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. The reference literature currently covers areas such as: pH 210 pH 212 • Water Treatment • Process • Swimming Pools • Agriculture • Food • Laboratory • Thermometry pH 211 pH 213 Microprocessor-based Bench pH/mV/°C Meters and many others. New reference material is constantly being added to the library.
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for a correct use of the instrument, as well as a precise idea of its versatility. If you need more technical information, do not hesitate to e-mail us at tech@hannainst.com. directives. These instruments are in compliance with CE DECLARATION OF CONFORMITY TABLE OF CONTENTS PRELIMINARY EXAMINATION .....................
WARRANTY PRELIMINARY EXAMINATION All Hanna Instruments meters are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrodes and the probes are guaranteed for a period of six months. This warranty is limited to repair or replacement free of charge. Damage due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
FUNCTIONAL DESCRIPTION pH 210 AND pH 211 Extension cables for screw-type electrodes (screw to BNC adapter) HI 7855/1 Extension cable 1m (3.3') long HI 7855/3 Extension cable 3m (9.9') long HI 7855 SERIES CABLE CONNECTORS CONNECTOR AND 3.0 mm (0.
HI 1413B Glass-body, single junction, flat tip, Viscolene, non-refillable, combination pH electrode. Use: surface measurement. SPECIFICATIONS p H 210 AND p H 211 12 mm 0.5" Range pH mV mV °C 0.00 to 14.00 ±399.9 (ISE, pH 211 only) ±1999 (ORP, pH 211 only) 0.0 to 100.0 Resolution pH mV mV °C 0.01 0.1 (ISE, pH 211 only) 1 (ORP, pH 211 only) 0.1 Accuracy (@20°C/68°F) pH mV mV °C ±0.01 ±0.2 mV (ISE, pH 211 only) ±1mV (ORP, pH 211 only) ±0.5 Typical EMC Deviation pH mV °C ±0.
FUNCTIONAL DESCRIPTION p H 212 AND p H 213 FC 100B Plastic-body (Kynar®), double junction, refillable, combination pH electrode. Use: general purpose for food industry. 12 mm 0.5" FC 100 120 mm 4.7" FC 200B Plastic-body (Kynar®), open junction, conic, Viscolene, non-refillable, combination pH electrode. Use: meat & cheese. 6 mm 0.25" FC 200 75 mm 2.95" FC 210B Glass-body, double junction, conic, Viscolene, non-refillable, combination pH electrode. Use: milk, yogurt. 12 mm 0.
HI 1330B Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: laboratory, vials. 5mm DIA 0.2" 5mm 0.2" HI 1330 120 mm 4.7" "S" VERSION HI 1331B Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: flasks. 8 mm 0.3" Range pH mV °C -2.00 to 16.00 / -2.000 to 16.000 ±999.9 (ISE and ORP, pH 213 only) 0.0 to 100.0 Resolution pH mV °C 0.01 / 0.001 0.1 (ISE and ORP, pH 213 only) 0.1 Accuracy (@20°C/68°F) pH mV °C ±0.002 & ±0.01 ±0.05% f.s.
OPERATIONAL GUIDE Power connection Plug the 12VDC adapter into the power supply socket (#10 on page 4; #11 on page 6). Note: pH 212 and pH 213 use an EEPROM to retain the pH, mV and temperature calibrations as well as the serial communication setting. The instrument will retain the data after a calibration or serial communication setting, even when it is unplugged. Note: Make sure the main line is protected by a fuse.
ACCESSORIES pH CALIBRATION SOLUTIONS HI 70004P pH 4.01 Buffer Sachets, 20mL, 25 pcs HI 70007P pH 7.01 Buffer Sachets, 20mL, 25 pcs HI 70010P pH 10.01 Buffer Sachets, 20mL, 25 pcs HI 7004L pH 4.01 Buffer Solution, 460 mL HI 7006L pH 6.86 Buffer Solution, 460 mL HI 7007L pH 7.01 Buffer Solution, 460 mL HI 7009L pH 9.18 Buffer Solution, 460 mL HI 7010L pH 10.01 Buffer Sol., 460 mL HI 8004L pH 4.01 Buffer Solution in FDA approved bottle, 460 mL HI 8006L pH 6.
ORP MEASUREMENTS (pH211 and pH213 only) TROUBLESHOOTING GUIDE An optional ORP electrode must be used to perform ORP measurements (see accessories). Oxidation-reduction potential (REDOX) measurements provide the quantification of the oxidizing or reducing power of the sample tested. To correctly perform a redox measurement, the surface of the ORP electrode must be clean and smooth. • Switch the meter on and press the "RANGE" key to enter the mV mode.
TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS The resistance of glass electrodes partially depends on the temperature. The lower the temperature, the higher the resistance. It takes longer time for the reading to stabilize if the resistance is higher. In addition, the response time will suffer to a greater degree at temperatures below 10 C.
CLEANING PROCEDURE • Press CAL. The "CAL" and " " indicators and the most common "7.01" buffer will be displayed on the secondary LCD. • If necessary, press "V C" or "W C" to select a different buffer value. • The "NOT READY" indication will blink on LCD until the reading has stabilized. • When the reading is stable, "READY" and "CFM" will blink. Press CFM to confirm the calibration. • If the reading is close to the selected buffer, the meter stores the reading.
For refillable electrodes: If the filling solution (electrolyte) is more than 2½ cm (1") below the fill hole, add HI 7082 or HI 8082 3.5M KCl Electrolyte Solution for double junction or HI 7071 or HI 8071 3.5M KCl+AgCl Electrolyte Solution for single junction electrodes. For a faster response, unscrew the fill hole screw during measurements. For AmpHel®electrodes: If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced.
TEMPERATURE CALIBRATION pH 212 AND pH 213 ONLY (for technical personnel only) ELECTRODE CONDITIONING & MAINTENANCE All the meters are factory calibrated for temperature. The temperature probes are interchangeable and no temperature calibration is needed when they are replaced. If, for any reason, the temperature measurements are inaccurate, temperature recalibration may be carried out. For an accurate recalibration, contact your dealer or the nearest Hanna Customer Service Center.
p H VALUES AT VARIOUS TEMPERATURES Temperature has an effect on pH. The calibration buffer solutions are affected by temperature changes to a lesser degree than normal solutions. During calibration the meter will automatically calibrate to the pH value corresponding to the measured or set temperature. TEMP °C °F 32 0 41 5 10 50 15 59 20 68 25 77 30 86 35 95 40 104 45 113 50 122 55 131 60 140 65 149 70 158 4.01 4.01 4.00 4.00 4.00 4.00 4.01 4.02 4.03 4.04 4.05 4.06 4.07 4.09 4.11 4.12 6.86 6.98 6.95 6.
INTERFACE WITH PC (p H 212 AND p H 213 ONLY) (pH Data transmission from the instrument to the PC is now much easier with the new HI 92000 Windows® compatible application software (optional). HI 92000 also offers an on-line help feature. Simply run your desired spread sheet and open the file downloaded by HI 92000. Data can be further elaborated with graphics, statistical analysis, etc. using the most diffused spread sheet programs (e.g. Excel©, Lotus 1-2-3©). To install HI 92000 you need a 3.