Instruction Manual HI 9033 • HI 9034 Waterproof Multi-Range EC and TDS Meters for Field Applications w w w . h a n n a i n s t .
PRELIMINARY EXAMINATION Dear Customer, Thank you for choosing a HANNA instruments® product. Please read this instruction manual carefully before using the instrument. It will provide you with the necessary information for a correct use of the instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. directives.
FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9033 1) 2) 3) 4) 5) 6) 7) 8) 9) 10) 11) Battery compartment Probe socket Liquid Crystal Display Measure unit Low battery indicator ON/OFF key 19.99 mS/cm range selection key Calibration trimmer 199.9 μS/cm range selection key 1999 μS/cm range selection key 199.9 mS/cm range selection key FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9034 1) 2) 3) 4) 5) 6) Battery compartment Probe socket Liquid Crystal Display Low battery indicator ON/OFF key 19.
FUNCTIONAL DESCRIPTION OF HI 76302W CONDUCTIVITY PROBE OPERATIONAL GUIDE INITIAL PREPARATION 1 Each meter is supplied complete with a 9 V battery. Remove the back cover, unwrap the battery and install it while paying attention to its polarity its (see "Battery Replacement" section). 3 4 2 1) 2) 3) 4) Watertight screened cable (1 m / 3.
TEMPERATURE EFFECT HI 76302W conductivity probe features a built-in temperature sensor that allows automatic compensation of the reading for any temperature differences. Wait for a few minutes for the temperature sensor to reach the thermal equilibrium with the test solution before taking measurements. If the temperature of conductivity probe and solution is drastically different, a longer time should be allowed.
• When the reading is stable, adjust the calibration trimmer to read the solution conductivity value @25°C. For example if using HI 7030, turn the trimmer to read 12.88 mS/cm @ 25°C. mS The calibration is now complete and the instrument is ready for use. All subsequent measurements will be compensated to 25°C (77°F). Note: Calibration can also be standardized to have readings compensated to 20°C (68°F), by simply adjusting the trimmer to the calibration solution value at 20°C (68°F).
CONDUCTIVITY VS. TEMPERATURE TDS VS. TEMPERATURE The conductivity of an aqueous solution is the measure of its ability to carry an electrical current by means of ionic motion. The conductivity increases with increasing temperature. In fact, it is affected by the type and number of ions in the solution, and by the viscosity of the solution itself. Both parameters are temperature dependent.
PROBE MAINTENANCE After evry series of measurements, rinse the probe with tap water. If a more thorough cleaning is required, remove the PVC sleeve and clean the probe with a cloth or a nonabrasive detergent. When reinserting the sleeve onto the probe, be sure that the sleeve is in the right direction with the four holes towards the cable end. After cleaning the probe, always re-calibrate the instrument. The probe body is made of PVC. For this reason it must never come into close contact with a heat source.
ACCESSORIES CONDUCTIVITY & TDS CALIBRATION SOLUTIONS HI 7030M HI 7030L HI 8030L HI 7031M HI 7031L HI 8031L HI 7033M HI 7033L HI 8033L HI 7034M HI 7034L HI 8034L HI 7035M HI 7035L HI 8035L HI 7039M HI 7039L HI 8039L HI 7032M HI 7032L HI 7036M HI 7036L 12880 μS/cm solution, 230 mL bottle 12880 μS/cm solution, 500 mL bottle 12880 μS/cm solution, 500 mL FDA bottle 1413 μS/cm solution, 230 mL bottle 1413 μS/cm solution, 500 mL bottle 1413 μS/cm solution, 500 mL FDA bottle 84 μS/cm solution, 230 mL bottle 84 μS/
CE DECLARATION OF CONFORMITY Recommendations for Users Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV equipment. The metal band at the end of the probe is sensitive to electrostatic discharges. Avoid touching this metal band at all times.
SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel. (21) 4584.2941 • Fax (21) 4584.2942 Japan: Tel. (03) 3258.9565 • Fax (03) 3258.9567 Korea: Tel. (02) 2278.5147 • Fax (02) 2264.1729 Malaysia: Tel. (603) 5638.9940 • Fax (603) 5638.9829 Singapore: Tel. 6296.