Product manual

Page 7For technical questions, please call 1-800-444-3353.SKU 95670
Measuring DC Voltage
Range Resolution Accuracy
200mV 100µV ±(0.5%+5)
2V 1mV
±(0.8%+5)20V 10mV
200V 100mV
1000V 1V ±(1.0%+5)
Input Impedance: 1MΩ for all ranges.
Overload Protection: 1000V DC/AC RMS
(For 200mV range: 250VAC)
1. Connect the red test lead to the
” Jack and the black
test lead to the “COM” Jack.
2. Set the Function/Range switch to the
desired range. If the voltage to be
measured is not known beforehand, set
the range switch to the highest range
and then turn it down range by range
until satisfactory resolution is obtained.
3. Connect the test leads to the
source or load to be measured.
4. Read the voltage value displayed
on the LCD along with the
polarity of the red test lead.
Measuring AC Voltage
Range Resolution Accuracy
200V 100mV ±(1.0%+5)
750V 1V ±(1.2%+5)
Input Impedance: approx. 450kΩ for all ranges
Frequency: 45Hz~450Hz
Indication: Average (RMS of sine wav
Overload Protection: 1000V DC/AC RMS
1. Connect the red test lead to the
” Jack and the black
test lead to the “COM” Jack.
2. Set the Function/Range switch to the
desired V~ range. If the voltage to be
measured is not known beforehand,
set the switch to the highest range and
then turn it down range by range until
satisfactory resolution is obtained.
3. Connect the test leads to the
source or load to be measured.
4. Read the voltage value
displayed on the LCD.