Manual

Hardy Bench Scales
3-6
Make sure the cable is not pinched and is clear
of the feet. The Bench Scale requires a recom
-
mended DC or AC excitation voltage of 5 volts
with a maximum of 15 volts. Wire the Scale cable
to an instrument or indicator following the color
code label found on the cable.
For HI BS200 and HI BS300 Scale Models:
+ Excitation Green
- Excitation Black
+ Sense Blue
- Sense Brown
+ Signal Red
- Signal White
Shield Bare
For HI BS400 Scale Models with C2 Electronic
Calibration:
+ Excitation Red
- Excitation Black
+ Signal Green
- Signal White
+ C2 Gray
- C2 Violet
Shield Yellow or Bare
WARNING USE CAUTION WHEN WIRING. MAKE SURE
TO DISCONNECT ALL POWER TO THE
SCALE. FAILURE TO REMOVE POWER CAN
RESULT IN SCALE OR INSTRUMENT DAM
-
AGE, DEGRADATION OF PERFORMANCE OR
PERSONAL INJURY
NOTE: Hardy Process Solutions provides drawings you
may download on the Hardy Bench Scales page
under the Docs & Programs tab. Drawings are
provided for each of the Hardy Bench Scale
models in .dxf, dwg, .pdf, and 3D formats. These
drawings include the wiring configurations.