User’s Guide Enviro™ Series Checkweighing Instrument © 2009 by Hardy Instruments, Inc.
Enviro™ Series Instrument Document Part Number 0596-0315-01 Revision B Hardy Instruments Inc. 9440 Carroll Park Drive, Suite 150 San Diego, CA 92121 Created 12/09 Created Document Revision B 12/09 Released Manual Disclaimer Every effort has been made to provide complete and accurate information in this manual.
Section 1: General Information Table of Contents SECTION 1 1.1. 1.2. SECTION 2 2.1. SECTION 3 3.1. 3.2. SECTION 4 4.1. 4.2. 4.3. 4.4. 4.5. SECTION 5 5.1 5.2 5.3 5.4 SECTION 6 6.1 6.2 6.3 SECTION 7 7.1. 7.2. 7.3. SECTION 8 12/09 GENERAL INFORMATION ..............................................................................................1 Introduction .........................................................................................................................1 Specifications ...........
8.1. 8.2 SECTION 9 Expanded Display Mode ..................................................................................................30 Scale Maintenance ...........................................................................................................30 8.2.1 Cleaning the Scale and Indicator ....................................................................30 PARTS LIST ...................................................................................................................
Section 1 General Information 1.1. INTRODUCTION The Enviro™ Series instrument is designed for light capacity, general purpose use. It is also designed for use in wash-down Environments. Features The Indicator has a capacity setting of up to six digits, using a six-digit amber LED display. Annunciators include, Net Weight, Units, Under, Accept, Over, and Accumulation. Checkweigh mode can store up to four Checkweigh recipes. 1.2. SPECIFICATIONS FEATURE DESCRIPTION Display 1.
Section 2: Service Policy Section 2 Service Policy Information 2.1. GENERAL SERVICE POLICY Customers are responsible to protect the equipment from accidental or malicious damage. Do not install or operate the equipment where the application requirements exceed the design specifications of the equipment. Do not make physical, electrical or program modifications to this equipment. Failure to comply with this policy voids all implied and/or written warranties.
Section 3: Installation Section 3 Installation 3.1. PRIOR TO INSTALLATION 1. Unpack and Check contents 2. Find best location 3.1.2 Unpacking Follow these guidelines when unpacking all equipment: Check in all components and accessories according to your order. Keep the shipping container and packing material for future use. Remove all components from their packing material, checking against the invoice that they are accounted for and not damaged.
Section 3: Installation IMPORTANT INSTALLATION NOTICE All load cells, load cell cables, and interconnecting cables used to connect all scale components shall be located a minimum of thirty-six (36”) inches distance away from all single and multiple phase high energy circuits and electric current carrying conductors. This includes digital weight indicators, junction boxes, sectional controllers, and power supplies.
Section 3: Installation 2. Remove the 120 VAC plug from the end of the power cord and attach a proper 220-240 VAC plug. Connect the green wire and the shield to the ground lug. Note: The original wiring connects brown to AC hot and blue to neutral. 3. Reattach all cables and replace the front cover assembly. Caution: Improper connections at J4 can cause catastrophic damage to the Indicator 3.2.
Section 3: Installation 3.2.4 Powering Up the Indicator The Indicator performs a warm-up cycle. The Indicator initiates a test, displaying numbers 1 to 8, and lights up all LED’s. The Program number and Revision Information displays.
Section 3: Installation Section 4 Programming Configuration 4.1. OVERALL STEPS Follow these steps to program the Enviro™ Series instrument. 1. Configure the Enviro™ Series Instrument operating parameters. 2. Calibrate the Enviro™ Series Instrument. 3. Set up the Enviro™ Series Instrument options. 4.2. INTRODUCTION The program group is shown and accessed in steps. Each program group may be entered and modified, or skipped to the next group.
Section 6: Serial I/O 4.3. KEYPAD FUNCTIONS, PROGRAMMING MODE. This chart shows what action will be taken when a front panel key is pressed in the programming mode. KEY ACTION 1 No function in programming mode. Exception, decrements selected digit when entering numeric data. 2 No function in programming mode. Exception exits to done prompt when entering numeric data. 3 No function in programming mode. Exception, moves flashing digit to far left position when entering numeric data.
Section 6: Serial I/O 4.5. PROGRAMMING STEPS Setup Menu 1. Press and hold the Program/Enter key or press S1 on the main PCB to display SEtuP. 2. Press the Program/Enter key to enter setup mode. 3. Press Zero to set the desired Programming Menu. Default selections are shown underlined. Programming time format: This will determine whether the clock is displayed and printed in 24 hour (military) or 12 hour (AM/PM) format. The display will indicate either 12 hour or 24 hour.
Section 6: Serial I/O P: Programming Port 1 Parity setting: The display will indicate p along with the current setting. Press over/under/next, units or tare to change the selection. Press Program/Enter to enter the selection. Available selections are: P nonE P odd P EvEn Busy: Programming Port 1 handshaking: The display will indicate xxbusy, where xx is either no, lo, or hi. Press over/under/next, units or tare to change the selection. Press Program/Enter to enter the selection.
Section 6: Serial I/O Opid: Include ID data with transmitted data: This determines whether or not to include the ID in the data string for Port 2. Display will indicate opid followed by the current selection of Y for yes to include, or n for no to exclude. Press over/under/next, units or tare to change the selection. Press Program/Enter to enter the selection. Int: Display brightness setting: This changes the intensity of the display brightness.
Section 6: Serial I/O Cust: Custom units. The display will indicate cust and the current selection, either y for enable or n for disable. Press over/under/next, units or tare to change the selection. Press Program/Enter to enter the selection. Note: When custom units are active, units LED indicators are off in the weigh mode. Azt: Programming the Automatic Zero Tracking band: (this feature will maintain zero when small amounts of material are placed on the scale, such as rain, snow, debris, etc.
Section 6: Serial I/O Press over/under/next, units or tare to change the selection. Press Program/Enter to enter the selection. Available selections are: OFF Tare is disabled. ON Tare is active. On-CLr Tare automatically clears when Gross weight returns to Zero Hl: Overload limit: The actual value at which the display goes to ol (represented as a percentage of capacity). The display will indicate hl followed by the current setting.
Section 6: Serial I/O X.xxupd: Display microvolts per division: This is a reference to display the current microvolts per division. The display will indicate x.xxupd where x.xx is a numeric value of the microvolts per division. Press any front panel pushbutton other than b/g / net to continue. Acc: Accumulate feature enabled: Enables or disables the accumulate feature. The display will indicate acc followed by the current setting of y for enable and n for disable.
Section 6: Serial I/O Section 5 Calibrating the Indicator 5.1 Introduction Calibration can be performed entirely from the front panel or locked out with security levels 2-5 or internal jumper JP1. Calibration can be performed with a single calibration point, or using up to four calibration points.
Section 6: Serial I/O 5.2 Single Point Calibration Procedure Press and hold the PROGRAM/ENTER key until “SEtUP is displayed. Or press internal switch S1 to display “SEtUP” Press the OVER/UNDER/NEXT key until “CAL” is displayed Press the PROGRAM/ENTER key to display security level “SL” Press OVER/UNDER/NEXT to select proper security level Press PROGRAM/ENTER Raw counts displayed No New Zero? Yes Press the ZERO key Scale zero’d.
Section 6: Serial I/O Displays “LOAd1”, then raw counts Apply test weights to scale. Press PROGRAM/ENTER key Displays correct weight. Note: UNITS LED blinking Press the 2 key or internal S1 Displays microvolts per division (x.xxuPd)* Press any front panel key to return to “CAL” prompt Press OVER/UNDER/NEXT to display “donE” Press PROGRAM/ENTER to display “SAvE” Press PROGRAM/ENTER to display “SAvEd” Scale returns to weigh mode *The “uPd” (microvolts per division) must be one (1.
Section 6: Serial I/O 5.3 Multipoint Calibration Press and hold the PROGRAM/ENTER key until “SEtUP is displayed. Or press internal switch S1 to display “SEtUP” Press the OVER/UNDER/NEXT key until “CAL” is displayed Press the PROGRAM/ENTER key to display security level “SL” Press OVER/UNDER/NEXT to select proper security level Press PROGRAM/ENTER Raw counts displayed No New Zero? Yes Press the ZERO key Scale zero’d.
Section 6: Serial I/O When display shows correct test weight press PROGRAM/ENTER Displays “LOAd1”, “LOAd2”, “LOAd3”, or “LOAd4” then raw counts Apply test weights to scale. Press PROGRAM/ENTER key Fourth span point entered Displays correct weight. Note: UNITS LED blinking PROGRAM/ENTER Press 2 key or internal switch S1 Displays microvolts per division (x.
Section 6: Serial I/O 5.4 Service Sealing Commercial Applications Scales used in a commercial application must be “placed in service” in accordance to local weights and measures guidelines. Four sealing screws are provided on the back cover to allow a lead seal to be installed if required. An audit trail is provided to track changes in programming and calibration. Access to the audit trail is as follows: 1. Press and hold b/g / net for approximately 10 seconds. 2. Indicator will display cfgctr.
Section 6: Serial I/O Section 6 Serial I/O 6.1 INTRODUCTION The Enviro™ Series Indicator is equipped with 2 serial output ports. Port 1 is a bidirectional rs232 port, and Port 2 is a bidirectional rs485 port. Port 1 can be configured to operate with a variety of printers, computer applications, and remote displays. 6.
Section 6: Serial I/O Remote display Please see the Enviro™ Series Printer Addendum for a list of suggested 3rd party remote displays and how to configure the Enviro™ Series check weighing instrument to work with them. Computer output The Enviro™ Series can be connected to a variety of computer systems utilizing the continuous or polled outputs. Connections to computers are done using the wiring chart provided earlier in this section.
Section 7: Operations Section 7 Operations 7.1. BASIC SCALE OPERATIONS Upon power up the Indicator performs a warm-up cycle. The Indicator initiates a test, displaying numbers 1 to 8, and lights up all LED’s. The Program number and Revision Information displays. The Indicator then displays the current weight on the scale 7.2. KEYPAD FUNCTIONS, WEIGH MODE KEY ACTION Recipe 1-4 Enables or disables programmed recipe in Checkweigh mode.
Section 7: Operations 3. Place container/object on scale (Tare weight). 4. Press the tare key. 5. Place material in container or add objects (Net weight). 6. Read the net weight on the display. Gross/Tare/Net Weighing 1. Press the B/G / Net key, if required, to set display to GR (gross). 2. Press the Zero key, if required, to set scale to “0.” 3. Place container/object on scale (Tare weight). 4. Press the Tare key. 5. Place material in container or add objects (Net weight). 6.
Section 7: Operations Port 1 set to Auto and Aout set to no. 1. After weight is applied, and if the weight is greater than the accumulate threshold, the indicator will flash all the green ACCEPT LEDs for one second showing that the weight has been stored as an accumulation. 2. If a print is to occur, press print and port 1 will transmit the following: “Weight nn = wwwwww uu GR/NT” nn is the number of the current accumulated weight in numerical order. wwwwww is the last accumulated weight.
Section 7: Operations 2. Weight on the scale must now be removed to less than .5 grads before the next accumulation can occur. 3. Repeat steps 1-3 to continue accumulations (up to 99). Displaying Accumulated Weight and Count. 1. Press and release Over/Under/Next The indicator will alternately display the total accumulated weight and the number of accumulations for two seconds each. The count is left justified followed by Acc, and the total weight is right justified. 2.
Section 7: Operations 3. Press Program/Enter to display Hold X, where X is either Y or N. 4. Press Over/Under/Next to display Hold Y if necessary. o 5. Note: Peak Hold function must not be used with Checkweighing. Press Program/Enter to display CSaved, and returns to weigh mode. Indicator is now in Peak Hold mode. To operate peak hold: 1. Press the 1 button to activate Peak Hold mode. The first three indicators for UNDER, ACCEPT and OVER will flash indicating Peak Hold is active.
Section 7: Operations 10. With the desired selection displayed, press Program/Enter to display units. With units displayed, one of the units legends will be lit. This is the weigh units the Checkweigh limits will be defined by in the following steps. 11. Press Over/Under/Next to scroll through the choices, including custom units. 12. With the desired units legend lit, press Program/Enter to display hand y, or hAnd n. This determines the method for defining the acceptance band.
Section 7: Operations 23. The entire under bar graph, the entire accept, and the first segment of the over bar graph will be illuminated. A previously programmed weight value is displayed with the leftmost digit flashing. This is the end of accept point. If required using the same procedure as identified in step 21 above, enter the correct value for the end of the accept range. When the correct end of accept value is displayed, press program/enter. 24.
Section 8: Service and Maintenance Section 8 Service & Maintenance 8.1. EXPANDED DISPLAY MODE The Enviro™ Series can be placed in expanded resolution mode to assist in troubleshooting and maintenance. Simply press S2 on the main PCB and the display indicator will immediately display the weight in 10X resolution. For example if the division size is programmed at 1lb, pressing S2 will display the weight in 0.1lb divisions. 8.2 SCALE MAINTENANCE 8.2.
Section 9: Parts Section 9 Parts List IP 69k Model ITEM PART # 1 29006 2 28117 3 29146 7 15131 8 11926 9 29008 10 30298 12 29007 16 11998 17 17545 18 15851 19 17534 20 12342 21 17503 22 17496 23 20963 24 20965 26 11495 28 14083 30 11191 31 11102 32 11124 35 11150 37 24366 39 29081 40 12609 42 25355 43 25356 44 12189 52 052-0086-01-0 12/09 QTY 1 1 1 1 AR 4 1 1 9 3 3 1 1 1 1 1 1 4 4 10 3 1 4 1 1 2 1 1 2 1 DESCRIPTION FRONT PANEL ASSY PCB ASSY, DISPLAY I/O CONTROLLER A1 PCB ASSY, SMALL BLOCK (NON-POTTED) A
Section 9: Parts IP 69k Reference Drawing 12/09 32 0596-0315-01 Rev B
Appendix I Keypad Reference Tb3 connections to keypad shown below. 01/10 33 0596-0315-01 Rev.
Enviro Series Instrument USER’S GUIDE Document Part Number 0596-0315-01 B