Instruction Manual
HI 2151 SERIES WEIGHT CONTROLLERS REMOTE I/O OPTION
5-4
Block Read
Command Number
2: Setpoint Relay
Parameter
BLOCK READ COMMAND NUMBER 2: Setpoint Relay Parameters
START
WORD DEFINITIONS: #WORDS WORD
Command number: A value of 2 (decimal)
bit 0 0
bit 1 1
bit 2 0
bit 3 0
bit 4 0
bit 5 0
bit 6 0
bit 7 0
Indicator Group 2 Status
bit 8 Weight currently displayed in pounds units
bit 9 Zero Track feature enabled
bit 10 Reserved for future use
bit 11 Current gross weight = 0
bit 12 Weight in motion, i.e. changing
bit 13 Gross weight currently displayed
bit 14 Net weight currently displayed
bit 15 Weight currently displayed in kilogram units
Note: A returned value of “99” (decimal) indicates an error.
10
Relay Status
bit 0 Relay #8 status (on/off)
bit 1 Relay #7 status (on/off)
bit 2 Relay #6 status (on/off)
bit 3 Relay #5 status (on/off)
bit 4 Relay #4 status (on/off)
bit 5 Relay #3 status (on/off)
bit 6 Relay #1 status (on/off)
bit 7 Relay #2 status (on/off)
bit 8 - 15 Setpoint description byte A (See Table 5-4 & 5-5)
11
bits 0-7 Setpoint description byte B (See Table 5-4 & 5-5)
bits 8-15 Setpoint description byte C (See Table 5-4 & 5-5)
12
Deadband value for setpoint #1
Deadband value for setpoint #2
Deadband value for setpoint #3
Deadband value for setpoint #4
Deadband value for setpoint #5
Deadband value for setpoint #6
Deadband value for setpoint #7
Deadband value for setpoint #8
2
2
2
2
2
2
2
2
3
5
7
9
11
13
15
17
Preact value for setpoint #1
Preact value for setpoint #2
Preact value for setpoint #3
Preact value for setpoint #4
Preact value for setpoint #5
Preact value for setpoint #6
Preact value for setpoint #7
Preact value for setpoint #8
2
2
2
2
2
2
2
2
19
21
23
25
27
29
31
33