DISPLAY TEST FAILURE SOLUTIONS
MOTION Indicates next parameter will be None
motion value.
3 Motion value calibrated in unit. Default is 3.
(example)
AVRAGE Indicates next display will be number of None
values set in the running average table
(1 - 200).
100 Number of values set up to be averaged. Default is 100.
(example)
LINCOR Indicates next display will be single point None
linear correction.
0 If 0, linear correction is not used. Number Default is 0.
(example) other than 0 indicates single-point calibrated
value.
AD OFF Indicates next display will be A to D None
Off Set.
28912 A to D Offset Value. None
ADGAIN Indicates next value will be A to D gain. None
828238 A to D Gain Value None
DIP 1 Indicates next display will be settings of If value is other than desired,
rear panel Config dipswitch S3 segments. reset dipswitch segment and
Settings represented by hexadecimal retest.
notation.
00 thru FF SWITCH
VA LUE ON OFF
POSIT
2 01 Cal toggle Cal toggle
4 02 Opt menu lock No opt menu lock
5 04 S.P. menu lock No S.P. menu lock
7 08 Tare, lb/kg, N/G, zero Keys not locked
lock
8 10 Zero track on Zero trk off
6 20 Spare Spare
3 40 Serial requests All serial I/O
only
1 80 P2 multidrop Not multidrop