Owner's manual

HI 2151/30WC MANUAL
8-38
Error Message Definition
Err 5 Not in Gross mode when trying to enter CAL menu.
Err 6 Invalid zero. Zero must be less than zero tolerance.
Err 7 Invalid grad size. Grad size must be greater than the full scale divided
by the range minus one valid grad size.
Err 8 Unable to enter Calibration Menu without toggling ReCal segment on
S3 or using the proper security code.
Err 9 Invalid zero. When the instrument is in NTEP mode, the zero must be
within the zero tolerance.
Err 10 Invalid motion tolerance. Motion must be greater than 0 and greater
then or equal to Grad Size.
Err 11 Negative gross weight during acquire tare. (NTEP only)
Err 12 All decades after decimal point must be active. (NTEP only)
Err 13 Full scale calibration of zero or negative value not allowed.
Err 14 Scale capacity of zero or at negative value not allowed or indicates
Err 15 Zero tolerance of zero or a negative value not allowed or indicates lin-
ear correlation value entered is not > 0 or < Span value.
Err 16 Tare greater than span. (NTEP only)
Err 17 Acquire tare cannot be used with zero gross weight. (NTEP only)
Err 18 Difference between zero and span is less then 100 counts in the positive
direction or indicates the linear correction point is not 100 counts > 0 or
100 counts < Span point.
Err 19 Occurs when configuring Serial Port 1 and entering a negative address.
Err 20 Option held when a print command was requested. Option hold active
on remote functions TB when print command from serial port
requested. (Serial Port only)
Err 24 Clock malfunction.
Err 25 Invalid total decimal point.
Err 26 Faulty time or date entry.
Err 40 Software Cal sticker value too small.
Err 41 Software Cal reference point weight out of range.
Err 42 Software Cal load cell overloaded.
Err 43 Software Cal range is zero error.
Err 44 Sensitivity out of range.
Err 45 No Software Cal in NBS range.
Err 50 Too many BCD Options.
Err 52 Too many Serial Options.
Err 54 Invalid serial or BCD format entry.
(_ _ _ _ _ _ P or _ _ _ _ _ _)