User Manual

7 CHAPTER 2
Specifications
CHAPTER 2: SPECIFICATIONS
About Chapter 2
Chapter 2 lists the specifications for the HI 3010 Filler/Dis-
penser. Specifications are listed for the standard instrument
and for instruments fitted with optional equipment. The
specifications listed are designed to assist in the installation,
operation and troubleshooting of the instrument. Service per-
sonnel should be familiar with this section before attempting
an installation or repair of the instrument.
Specifications for a Standard Instrument
Update Rate:
55 updates per second
Resolution:
Displayed: 1:985,000 (3 mV/V load cells)
Internal 1:1,048,5761
Excitation Voltage:
•5 VDC
Drives up to 8 350 ohm load cells
Averages:
1 to 250 - Sliding, User Selectable in Sin-
gle Unit Increments
Input:
Up to eight (8) 350 ohm Full Wheatstone
Bridge, Strain Gauge Load Sensors/Cells
(5 volt excitation) on one vessel.
Signal Voltage Range ± 0-120mV/V (±
600 mV)
Display:
4 line x 20 character backlit LCD
5 x 7 dot matrix
Display Increments (Graduations):
1,2,5,10,20,50,100,200,500 user selectable
Corresponding weight is dependent on the
decimal point location.
Standard Electronic AC Relays:
Wire Size: 12 AWG Maximum
Maximum Switch Current: .5 Amps
Maximum Switch Power: 120 Watts
Maximum Switch Voltage: .5 Amps @
240 VAC
Single Cycle Surge: 85 Amps (Peak)
Standard Electronic DC Relays:
Wire Size: 12 AWG Maximum
Maximum Switch Current: .5 Amps
Maximum Switch Voltage: .5 Amps @ 60
VDC
Switch Voltage: 5-60 VDC
1 second surge: 5 Amps
Non-Linearity:
0.0015% of Full Scale
WAVERSAVER
®
:
•Off
•7.5 Hz
•3.5 Hz
•1.0 Hz
•0.5 Hz
0.25 Hz
Calibration Techniques:
•C2
®
- Second Generation: Electronic
Traditional - Calibration with test weights
Standard Interfaces:
Ethernet - 10/100 Base T; embedded
server
DeviceNet - Master Scanner
Serial RS 232 - simplex to Printer.
Excitation Monitor:
Current less than 10%
Power and Utility Requirements:
Voltage - Universal Power Supply (50/60 Hz)
120-240 VAC ± 10%
12 - 24 VDC
Frequency
•47-63 Hz