User guide

Instrument Configuration
47
Calibration Menu
C2 eCAL
o Sensitivity
oGravity
oRef Wt
o Do C2 Calibration
oCal Tol
oNum Dev
Hard Cal
oCal Tol
o Sensitivity
oCal Lo Wt
oDo Cal Lo
oSpan Wt
oDo Cal Hi
Cal Date
C2 (eCal) Commands and Parameters
Cal Tolerance Parameter
The value you enter for Calibration Tolerance sets the amount of deviation to allow during
the calibration process. This value must be greater than or equal to the base motion value
and/or the Graduation Sizes.
The base motion value can be calculated by using the following formula:
Base Motion value = (Total Load Cell Capacity x 0.0003)
We recommend three graduation sizes.
RANGE: .000001 - 999999 (default 10)
On the Web page, enter the value in the text box provided.