Instrument Configuration
69
Span Weight Parameter
The Span Weight value is the weight of the object that is being placed on the scale to
set the “High” calibration point with respect to the “Low” calibration point set using
the Cal Lo Weight parameter.
On the Web page, select the calibration menu and left click inside the Span Weight
text field and enter the load cell sensitivity value.
On the
HI 6500 series instrument display
, select the calibration menu and use the UP or
DOWN button to select the Hard Cal submenu, then select the Span weight menu item. The
current Span Weight value will be displayed, press the ENTER or the RIGHT button if this
value needs to be modified.
Cal Date Parameter
The Cal Date parameter is used to record the date when the last calibration was completed
on the HI 6500 series instrument. The format is
mm/dd/yyyy
On the
Web page
, select the calibration menu and left click inside the Cal Date text field
and enter the date the calibration was completed.
On the
HI 6500 series instrument display
, select the calibration menu and use the UP or
DOWN button to select the Cal Date submenu. The last Cal Date value will be displayed,
press the ENTER or the RIGHT button if this date needs to be modified.