User Manual

Calibration:
Sectionusedtocalibratetheinstrumenttothesensor(s)used.
C2ElectronicCalibration
1) Removeallweightfromthescale.
2) PressDoC2Cal.
3) Placeaverificationweightonthescaletoensurecalibrationis
successful.
HardCal‐
1) InputtheSpanWeight.Thespanweightshouldbeabout80%of
thescalecapacity.
2) InputCalLowWeight.Withnothingonthescale,theCalLow
Weightis0,with5lbsonthescale,theCalLowWeightwouldbe5.
3) PressDo
CalLowtosetthelowpointonthecalibrationcurve.
4) PlacetheSpanWeightontothescale.
5) PressDoCalHi.
RefertoUser’sGuideifadjustmentstoothercalibrationrelated
parametersarerequired.
Communication:
Sectionusedforsettingupcom municationstotheinstrument.
EthernetTCP/IP
1)
ConnectanEthernetcablebetweentheinstrument’sRJ45
connectorandaPC,thenpoweruptheinstrument.Acrossover
cableisnotrequired,butcanbeused.
2) Theinstrumentwillautonegotiatesettingssuitableforavarietyof
operatingsystemsandnetworkconfigurations.Waitapproximately
15secondsand
checktoseeifthegreenlightoftheEthernetportis
onandblinkingifso,skiptostep8youareconnected.
3) IfthegreenlightontheEthernetportdoesnotlightup,checkthat
theunit’sEnableDHCPisturnedoffandfollowsteps4through9.
4) Usingtheunit’sFixedIP address(thedefaultis192.168.000.100),
thePCmustnowbeassignedauniqueIPaddress(forexample,
192.168.000.101)