Running the Stability Test from the Web Interface
A Stability Test column on the IT test results display (see picture above) shows PASS or
FAIL for each load sensor.
We obtained the data by running the IT test, as described above.
Stability test is in the Diagnostics menu. Select the Stability Test hyperlink and wait a few
seconds.
The results will not be specific to each load sensor, as in the IT test. They will display an
average reading and
PASS/FAIL for the system as a whole.If equipped with an IT summing
junction box there will be a selection to run
the Weight and voltage test. That test will
display the individual load cell and loop back test results.
Running the Stability Test from the Optional Front Panel
1) From the Diagnostic menu, select Stability Test with use the down arrow and press
Enter. The HI 6600 runs
the test and shows the results.
2) To see the AD/Raw Count and the A/D Filtered select them from the menu.
3) Press Exit to return to the Test Menu.
Stability:
Mean
Variation
Results
mV/V
RTZ
Gross = 0.9lb
RAW
32918
0.17
OK
0.0184
Pass
Waversave
r
32993
0.00
OK
AVERTISSEMENT – Ne pas effectuer le test de stabilité lors de la production.
Ces tests peuvent résulter à des lectures incorrectes.
WARNING - D
O NOT PERFORM THE
S
TABILITY
T
EST DURING PRODUCTION
. T
HE
TEST ACTIVITIES CAN CAUSE INCORRECT READINGS
.