VuLink CI Operator’s Manual In-Situ, Inc | 221 E.
Contents Symbols ................................................................................................................................................................................................. 5 Safety Information............................................................................................................................................................................. 6 Required Components......................................................................................
Controls ...............................................................................................................................................................................................25 Specifications ....................................................................................................................................................................................26 1-970-498-1500 3 www.in-situ.
1 Topic header Symbols Important Symbols in This Manual The exclamation point calls your attention to a requirement, safety issue, or important action that should not be overlooked. A check mark highlights a tip or feature. Important Symbols on the Product Caution This symbol indicates critical safety information. Ignoring text that accompanies this symbol could result in injury or death due to improper handling.
Safety Information Using VuLink Correctly Read these instructions carefully before using VuLink. Don’t use VuLink in any manner not specified in the manual or quickstart guide. Follow all safety warnings. Installing and Replacing Batteries Never mix old and new batteries, or Lithium and alkaline batteries. Make sure all three batteries are installed in the same orientation. Use only In-Situ recommended Lithium batteries for longest battery life.
Required Components 1 Cable Rugged Twist-Lock Cable Connects VuLink to an Aqua TROLL, Baro TROLL, Level TROLL, or Rugged TROLL instrument. Vented or non-vented. 2 Instrument Aqua TROLL Baro TROLL Level TROLL Rugged TROLL 3 Software HydroVu Software VuSitu Mobile App View data, manage instruments, create alarms, and modify VuLink settings in your browser. Communicate with VuLink on any Bluetooth-enabled mobile device and the VuSitu mobile app.
Accessories The following accessories are available for VuLink. Cellular Antenna Part #: 0043630 The cellular antenna permits strong cellular network connectivity. Rugged Cable Splitter Part #: 0095500 (vented) Part #: 0085840 (non-vented) With the Rugged Cable Splitter, you can connect as many as 8 instruments to VuLink. Load-Bearing Universal Adapter Part #: 0101000 To attach pulse instruments and devices that don’t have a Twist-Lock connector, use the Load-Bearing Universal Adapter.
1 Topic header How it Works Using VuLink in any manner not specified by the manufacturer (In-Situ) may impair the device’s built-in protections. 1 Carabiner VuLink hangs from the top of a well with the detachable carabiner. 2 VuLink Telemetry Device VuLink provides power to monitoring instruments, transmits data to the cloud, and notifies you when maintenance is required. 3 Instrument An In-Situ instrument measures water quality and/or water level. 1-970-498-1500 8 www.in-situ.
VuLink CI 1 Create a HydroVu account. Visit hydrovu.com and create an account. 3 Scan the QR code on your VuLink. or Open your web camera and scan the QR code on your device, or type the registration code into the provided field. In-Situ, Inc | 221 E. Lincoln Ave | Fort Collins, CO 80524 | USA 2 Go to the telemetry page. Click the telemetry page link in the menu on the left side of the page. Then click Add a VuLink. 4 Attach the antenna and connect an instrument.
Setting Up Vulink VuLink works with Aqua TROLL, Level TROLL, Baro TROLL, and Rugged TROLL instruments. Follow the steps below to connect an instrument and start transmitting data. 5 Attach the antenna and remove the battery pull-tab. Make sure the antenna is properly attached before proceeding. Remove the battery cover by twisting it counter-clockwise and pulling down. Remove the yellow tab to allow current to flow through the batteries. Replace the cover. Press the button. All LEDs turn on.
Adding Instruments 6 Connect an Instrument cclilicckk Align the flat edge of the connector with the flat edge inside the cable. Twist the cable until it clicks into the secure postion. Connect the cable to your instrument. 7 Add another instrument and press the button (optional) Add as many as 8 instruments to VuLink, repeating the steps below for each one. When you’re ready to deploy, use one or more Rugged Cable Splitters to connect multiple instruments. Disconnect instrument.
Blinking red and green A blinking red and green LED indicates a problem with the batteries. Do not deploy the VuLink in this condition. Check batteries and reinstall as necessary.
Logging With VuLink VuLink logs get uploaded to the cloud; instrument logs do not. Be sure to understand the differences between these log types before deploying VuLink. VuLink Logs 1 2 Instrument Logs Program a log into VuLink with the VuSitu mobile app. Program a 1 log into the instrument with VuSitu. 2 See your data in HydroVu. Download log files with VuSitu. Pressing the VuLink button for 5 seconds or longer will stop the active VuLink log.
Using HydroVu Use HydroVu to create logs, configure alarms, and modify VuLink’s settings. Sidebar Menu Options/HydroVu Pages View your data See Locations Create and edit dashboards Configure telemetry devices Manage alarms Upload data Set up notifications Edit calculated parameters Manage users 1-970-498-1500 14 www.in-situ.
Using VuLink With VuSitu After connecting to your VuLink with VuSitu, the app always displays the Connected Telemetry Device screen at launch. You can access all features of the app from this screen. Connected Telemetry Device Screen Access help information. Access menu. View and make changes to instruments connected to a VuLink. Specifiy a destination for uploads, like HydroVu or an FTP server. 1-970-498-1500 Create and modify logs. Adjust VuLink and instrument settings. 15 www.in-situ.
Creating Alarms VuLink Alarms VuLink alarms send a notification when VuLink batteries run low, a VuLink parameter crosses a specified threshold, or other events occur at a remote monitoring site. Tap All Settings from the menu. Tap Real-Time Alarms. To create an alarm, tap Add a real-time alarm. Select the parameter that should trigger the alarm and set the limits.
Managing Alarm Responses Alarms can send a notification by SMS or through the cloud when an alarm threshold is crossed. Message and data rates may apply. Tap All Settings from the menu. Tap Manage Alarm Responses. Select cloud notifications or SMS notifications. Enter SMS phone number(s). Select Apply and Test. Confirm that the test message was received. 1-970-498-1500 17 www.in-situ.
Using VuLink with an FTP Server You can configure VuLink to upload data to an FTP server via VuSitu. Have your FTP hostname, path, port, username, and password ready before getting started. a b Tap All Settings. c Select Telemetry Cloud Service. e f Enter your FTP VuLink tests the credentials. Then tap Test connection to the server. & Save. d Tap the radio button next to FTP. g The app displays the test results. Connect to VuLink with the VuSitu mobile app.
Read the pop-up If the upload is message about data successful, VuSitu charges. Tap Send Data if displays a confirmation. you wish to continue. 1-970-498-1500 19 www.in-situ.
Connecting a Pulse Instrument to VuLink VuLink can upload data from a pulse instrument to HydroVu or an FTP server. Required Components • Rugged Twist-Lock Cable with one stripped-and-tinned end • Load-Bearing Universal Adapter (LBUA) • Cable (from LBUA to pulse instrument) • Pulse instrument • VuLink Wiring with the Load-Bearing Universal Adapter (LBUA) To VuLink To pulse instrument Attach the Twist-Lock Connect the brown and end of a Rugged Cable to black wires at the other VuLink.
Setting Up a Pulse Instrument with VuSitu Add rain gauges and other pulse devices to VuLink with VuSitu pulse configuration. 1 Add the instrument to VuLink. Launch VuSitu and connect to VuLink. Tap the Connected Instruments dropdown. Connect the instrument to VuLink with a cable. Press the Add button in VuSitu. VuSitu displays a confirmation message. Press OK to dismiss it. Tap Add New. Select the Pulse Instrument radio button. Select low or high frequency.
Tap the gear icon next to the unit field to select a unit. Enter the value of one pulse in the selected units. Tap Save. VuSitu displays a “Saving” message. High frequency: Select High Frequency and choose a parameter and unit. Enter a minimum and maximum value. Enter a minimum and maximum frequency in hertz. Tap Save to confirm the pulse instrument configuration. VuSitu displays a “Saving” message.
Understanding VuLink SIM Cards External SIM Card VuLink attempts to use an external SIM card for all communications if one is present. If communication via the external SIM fails, VuLink uses the built-in SIM instead. Built-in SIM If an external SIM card isn’t present, VuLink uses its built-in SIM for all communication. Updating VuLink The automatic update option ensures that VuLink always has the current firmware. 1-970-498-1500 23 www.in-situ.
Controls LEDs on VuLink’s control panel indicate the device’s status. 8 1 7 2 6 3 5 1 Carabiner 2 Bluetooth status 3 Battery status 4 4 Connection status 5 Network connection status 6 Cloud connection status 7 Antenna 8 Power 1-970-498-1500 24 www.in-situ.
Specifications Battery Operation Time (24 hour reporting, LiMnO2) Operation Time (24 hour reporting, Alkaline) Clock Accuracy Network Type Bands Protocols Data Provider Antenna GPS File Format (nonHydroVu) 3 x D cell (1.5V - 3.
Storage Temperature Operating Temperature Ingress Protection Protocols Connectors Simultaneous Connections Venting Barometric Compensation Barometer Accuracy -20°C to 60°C -20°C to 50°C (Li-SOCl2/Li-MnO2), 5°C - 40°C (Alkaline) Device: IP68 System: Up to IP68 per antenna specification Modbus over RS-485, SDI-12, Pulse low/high frequencies (max 40 khz) 1 In-Situ Twistlock (supports multiple instruments via Rugged Cable Splitter, TROLL Net Hub, or Load-Bearing Universal Adapter) Up to 8 instruments (total