Use Instructions
Table Of Contents
- Chapter 1: Introduction and Product Safety Information
- Chapter 2: Product Introduction
- Chapter 3: Headset and Application Use
- 3.1 Minimum System Requirements
- 3.2 Cybersecurity
- 3.3 First Time Setup
- 3.4 App Navigation
- 3.5 Device Association
- 3.6 Subject Information Management
- 3.7 Cleaning
- 3.8 Test
- 3.9 Help Screens
- 3.10 Headset Management
- 3.11 Firmware Upgrade
- 3.12 Charging
- 3.13 LED Indicators
- 3.14 Account Settings and Logout
- 3.15 Other Settings
- Chapter 4: Troubleshooting
- Chapter 1: NOTE: The Headset when powered off enters a sleep mode, in which the device will draw less power but still drain the battery overtime if not placed on charge. It is recommended that if the device is to be used within 2 weeks of powering off...
- Chapter 5: Legal
Instructions for Use
QSL01-001-rev8.0-A
HS01-019
Page 19 of 42
1. DISINFECT
2. DRY
3. RINSE & WIPE
4. DRY
3.7.1 Disinfect
a)
Ensure gloves are worn.
b)
Remove used foam cylinders from the Headset.
c)
Ensure charging port cover is closed (see Figure 2-1 C, page 9).
d)
With a disinfectant wipe (for example, Caviwipes), wipe all surfaces of the Headset to remove visible
soil for a minimum of 1 minute. Pay extra attention to the seams, crevices and gaps of the Headset, by
cleaning these with the following method:
• Place a cotton swab on a disinfectant wipe and fold the wipe down around the swab, as
displayed in below.
• Using the cotton swab wrapped in the disinfectant wipe, wipe the seams, crevices and gaps
of the Headset and the following CRITICAL AREAs:
- Underside of the USB Port Cover and pay particular attention to the inner edges as
seen below.
- Inside of each Electrode Frustum – create full circular motions with the cotton swab
inside each frustum.
- Inside crevice of the Visor Cushion, identified in red in the figure below.