Use Instructions

Table Of Contents
Instructions for Use
QSL01-001-rev8.0-A
HS01-019
Page 25 of 42
g)
Impedance Test
Headset:
Good skin contact is indicated by green circles. Poor skin contact is
indicated by amber squares. Adjust Headset by gently shifting the
rear portion of the Headset until any amber squares are replaced by
green circles.
If any of the indicators remain amber squares, remove Headset from
subject, re-dampen foam cylinders and re-fit the Headset on
subject’s head as per Step f).
When all of the indicators are green circles, inform the subject of the
following:
The test is about to begin,
There will be two rounds of flashing lights for 30
seconds each, with a rest in between
Eyes are to remain open at all times (blinking is OK)
Avoid movement (e.g. looking around or away) of eyes
The test may be stopped at any time.
Smartphone:
When all three indicators are represented by green circles and the
subject is ready, select ‘BEGIN TEST’ to start the first test.
h)
After the first test begins, a progress graph will be displayed during
testing.
To cancel the test at any time, select ‘CANCEL’.
i)
At 100% (both ‘Test Progress’ and ‘Processing’ Results bar graph) the first test will stop.