Use Instructions

Table Of Contents
Instructions for Use
QSL01-001-rev8.0-A
HS01-019
Page 37 of 42
Headset Suddenly Off
The Headset is designed to turn off after 5 minutes if no test has been performed. Please power on the device to
continue use.
Flashing Amber LEDs Upon Powering the Headset
If the Headset flashes amber while it is turned on, this means the battery is insufficient for a test. Please power
off the device and charge it with the provided charger for at least 30 minutes.
Chapter 5: NOTE: The Headset when powered off enters a sleep mode, in which
the device will draw less power but still drain the battery overtime if not placed on
charge. It is recommended that if the device is to be used within 2 weeks of
powering off, that it be placed on charge to have sufficient battery for use once
powered back on. Legal
EU Waste Electrical and Electronic Equipment (WEEE) Directive
The symbol on the left means used electrical and electronic equipment (WEEE) should not be mixed
with general household waste. For proper treatment, recovery and recycling, please take this
product(s) to designated collection points where it will be accepted free of charge. Alternatively, in
some countries, you may be able to return your product(s) to your local retailer upon purchase of an
equivalent new product.
Disposing of this product correctly will help save valuable resources and prevent any potential
negative effects on human health and the environment, which could otherwise arise from
inappropriate waste handling.
Please contact your local authority for further details of your nearest designated collection point.
Penalties may be applicable for incorrect disposal of this waste, in accordance with your national
legislation.
Electromagnetic Compatibility
This device is intended to be used in the electromagnetic environment specified below:
Emissions Level
Compliance
Guidance
RF emissions
(CISPR 11)
Group 1
Group 1 equipment is equipment which does not use high
powered radiofrequency energy. Therefore, it is unlikely
to cause any interference with nearby electronics
equipment.
Class B
Class B equipment is equipment suitable for use in
establishments directly connected to a low voltage power
supply network. For example, in general practice clinics.
Electrostatic
discharge immunity
61000-4-2
±15 kV
This equipment is resistant to electrostatic discharge.
However, use in low humidity environments (<30% RH)
should be avoided if possible.
Radiated RF IEC
61000-4-3
80 MHz to
2.7 GHz
3 V/m
The NUROCHEK-I System complies with all applicable
electromagnetic compatibility requirements (EMC)
according to IEC60601-1-2:2014, for residential,
commercial and light industry environments. Portable
and mobile RF communications equipment should be
used no closer to any part of the machine, including
cables, than the recommended separation distance.
Should you suspect that the device performance (e.g.
connection quality) is affected by other equipment:
Reorient or relocate the device receiving the
interference.