BCS-G6 Series Body Composition Scales User Instructions PELSTAR, LLC 9500 West 55th St. McCook, IL 60525-7110 USA www.homscales.com © Pelstar, LLC 2015 Rev 3.
Table of Contents Section I - Installation ..................................................................................................................... 5 Caution and Warnings................................................................................................................. 5 Scale Specifications ..................................................................................................................... 6 Receiving and Inspection ...............................................
Patient ID Option ...................................................................................................................... 26 Disabling/Enabling the Calorie Option...................................................................................... 26 Section VII – Maintenance ............................................................................................................ 27 Serial Number and other Technical Information .............................................................
Section I - Installation Caution and Warnings • • • • • • • • • • • • • • • • Individuals who have internally implanted medical devices, such as Pacemakers, should not use this equipment due to the risk of malfunction to the device that may be caused by the weak electrical current. To prevent injury and damage to your scale, please follow these instructions very carefully. Assemble and operate the scale per the enclosed user instructions.
Scale Specifications Capacity and Resolution 1000 lb x 0.1 lb Power Requirements 110Vac/220Vac to 6VDC/2A Environmental Operating temperatures: 0°C to 50°C Storage temperatures: -15°C to 65°C) Humidity: 70% or less up to 40°C non-condensing Physical Dimensions Platform Length: 20” Width: 18” Height: 2.25” Scale Length: 23.5” Width: 18” Height: 41.5” Weight: 27 lbs. Receiving and Inspection Each BCS-G6 scale is shipped disassembled in one carton.
Scale Parts 7
Detail ATTENTION - WARNING Power Status LED is an indicator that shows the status of scale’s internal computer. To avoid damage to the operating system, never disconnect the power supply from the scale while the LED is ON or FLASHING. Switch the scale off with the ON/OFF switch and wait for the LED to turn off before disconnecting the power supply from the scale.
Assembly Step 1. Remove the base from the box and install the four feet. NOTE: The feet sub-assembly, come pre-assembled. There are 3 components to each foot assembly: • Foot • Space • Nut DO NOT DISASSEMBLE FOOT Removing the nut and/or spacer from the foot assembly will cause the scale to malfunction.
Step 2. Connect the plug from the stem to the receptacle on the base. To properly connect, the notches on the plug and receptacle must align. Step 3. Align the 4 studs from the stem to the 4 holes on the base and insert.
Step 4. While holding the stem and studs in place, carefully tip the scale at an angle to view the bottom of the base. Insert the 4 lock washers and tighten the butterfly nuts to the 4 studs.
Connecting Scale to Printer (USB Setup) The BCS-G6 is designed to work with several HP printers. Before purchasing a printer, please check the list of compatible printers. (View the list online at www.homscales.com) 1. Before making connections, make sure that the scale is off. 2. Turn on the printer and follow any first time power up and alignment procedures. 3.
Installing the Wireless Device (optional) If you have purchased the wireless option for your BCS-G6, in addition to your scale and hardware, you will also receive the following: a. b. c. d. TP Link Antenna Model TL-WN822N Antenna Bracket Mini USB Cable SD Card with Wireless Drivers Make sure that the scale is off, then use the following steps to install the antenna to your scale. (See illustrations on the following page.) Remove the film from the double-side type on the back side of the bracket.
Connecting Scale to Wireless Printer If you have purchased a BCS-G6 system with the wireless option, make sure to have the wireless device installed. Before purchasing a wireless printer, please check the list of compatible “Wireless Direct” printers online at www.homscales.com. To set up the printer to receive data from the scale: 1. 2. 3. 4. 5. Turn on the printer and follow any first time power up and alignment procedures. Turn on the “Wireless Direct” function on the printer.
Powering up Before plugging the power supply into the wall, verify that the ON/OFF switch located on upper part of the stem is in the OFF position. Connect the power supply to the wall socket and then to the scale. Once the connection is made, flip the switch to the ON position. Attention: To avoid failure, never disconnect the power supply from the scale while the Power Status LED is ON or FLASHING.
When the BCS-G6 turns on, the screen will briefly display a start-up message which includes the Serial Number and other technical information. The startup will be followed by a synchronization countdown. Once the display goes blank, the scale is ready to be used.
Section II - Operations How to Take a Measurement 1. Have the patient remove their shoes and socks and stand on the scale1. 2. Ensure the patient remains still while the scale measures their weight. 3. When prompted, enter the Age. On BCS-G62 Duo scales only: For children (4 to 17y11m) the age must be entered in Years and Months. From 4 to 9 you will need to press enter after the year is entered, from 10 to 17 the scale will automatically skip to the month field. For Adults 18 to 99, no months are required.
7. To print the results, select graphic or plain and press ENTER. (See Section IV - Printouts for details on the print types). How to Recall the Last Measurement The BCS-G6 allows the user to recall the last measurement and reprint it. To recall the last results: 1. Press the RCL button on the display panel. The data will be displayed. The display is be cleared by pressing the CLR button. 2. To print the results, select graphic or plain and press ENTER.
Section III – Understanding Results and Norms In order to determine a discernable change in body composition a 3 % change in body weight is recommended.2 Body Weight: Indicates the gross total body weight. % Body Fat: Total Percent of Body Fat.
Body Mass Index: BMI is a value derived from the mass (weight) and height of an individual. The BMI is defined as the body mass divided by the square of the body height. % Body Water: Proper hydration norms are based on % body fat4. Note: G6-Duo does not display %Body Water in Pediatric Mode.
Fat Free Mass: Fat Free Mass refers to the portion of the body composed of muscle, bones, organs and fluids. Caloric Burn Rates: Caloric Burn Rates are determined by adding the REE calories to various levels of physical activity. For your own calculations, remember 3500 calories make one pound of fat. Visceral Fat: Visceral fat is body fat that is stored within the abdominal cavity and is therefore stored around a number of important internal organs such as the liver, pancreas and intestines.
Section IV - Printouts When the results are displayed, the BCS-G6 allows you to select the print type, Plain or Graphic. Plain Printouts The Plain printout is a single page containing all the results from the measurements and calculations. Programming Number of Copies The user can pre-define the number of copies to print. The default value is 1. To change the value: 1. Without stepping on the scale, Press # 3 to access Settings Menu – Menu Password (PW): 7538 2. When prompted [MENU ?], press Plain. 3.
Section V - Special Modes Weight Only Mode The BCS-G6 allows you to take weight only measurements. To enable this measurement mode: Press Weight Only before stepping onto the scale. The scale will display WGT ON. Step on the scale and press Enter. The display will count down from 5 to 1 and display your weight.
Section VI – Customizing Lock Option The Lock option allows you to “lock” the BCS-G6 and prevent it from being used by unauthorized people. When the BCS-G6 is locked, you will be required to enter an authorization ID before each test. To enable the Lock feature: 1. Press 3 to access Settings Menu – Menu Password (PW): 7538 2. When prompted [MENU ?], press # 9. The screen will display: [PIN? Y/N?] 3. Press the YES key to enable the Lock Pin Option, NO to cancel. The default authorization ID number is 1234.
Physician ID The BCS-G6 allows the user to set a physician’s ID. The ID can be a number up to 8 digits and if set will be printed on the printouts under “Prepared By:” To set the Physician ID: 1. Press 5 before stepping on the scale. The screen will display: [ID 0000 0000] 2. Enter the preferred ID, up to 8 digits and press ENTER. The Physician ID can be changed or reset at any time. Patient ID Option The Patient ID option gives you the choice of using a Patient ID with every test.
Section VII – Maintenance The following pages provide instructions for maintenance, cleaning, calibration, and troubleshooting for the BCS-G6 scale. Caution: Before first use, or after long periods of non-use, check the scale for proper operation and function. If the scale does not operate correctly, refer to qualified service personnel. 1. Check overall appearance of the total scale for any obvious damage, wear, and tear. 2. Inspect the AC adapter for cord cracking or fraying, or for broken/bent prongs.
Calibration For Calibration instructions, please contact Health o meter® Professional Scales Customer Service at 1-800-815-6615. Zero Calibration Press 0 when scale is idle to reset any zero offset. Cleaning Proper care and cleaning is essential to ensure a long life of accurate and effective operation. Caution: Disconnect the scale from the power supply prior to cleaning the unit. ATTENTION - WARNING Power Status LED is an indicator that shows the status of scale’s internal computer.
Troubleshooting Issue Unable to print 7 Cause Solution Incompatible printer Verify that the printer being used is on the list of compatible printers. Check if ink cartridges are low and need to be replaced. Make sure that the ink cartridges are authentic HP. Some printers do not work well with refurbished or third party cartridges. Verify that there is paper in the correct paper tray. Verify that the USB cable is firmly connected to both the scale and printer. Replace the USB Cable. Turn off scale.
Section VIII – Compatible Printers Please see enclosed list of compatible printers or check our web site at www.homscales.com to find the most current list of printers that will work with your BodyComp Scale. For additional information, call 1-800-815-6615 Visit our website at www.homscales.