Instructions / Assembly

16
TAKING YOUR BLOOD PRESSURE READING (continued)
NOTE: The default setting of this monitor is to inflate the cuff to approximately 190 mmHg.
After three readings, the cuff will inflate to approximately 40 mmHg higher than the average
of the last three systolic test reading averages stored in memory. If the system detects that
the cuff needs additional pressure, it will automatically re-inflate to approximately 40 mmHg
higher than the last pressure level.
RECALLING MEASUREMENTS IN MEMORY
You can recall up to 120 total measurements in memory (60 per bank).
1. First, select the correct User bank, press the SET button. The User
symbol will begin to blink. Then press the MEMORY button to switch
between User 1 & 2. Once the desired User is selected, press and
hold the POWER button until the unit powers OFF.
2. Press and release the MEMORY button. It will display the average of
the last three readings, Fig. 2.
3. Press the MEMORY button again to scroll through the saved
measurements.
NOTE: When the number of readings exceeds 60 per bank, the oldest data
will be replaced with the new record.
To erase all stored measurements:
1. Press and release the MEMORY button to enter into the memory bank
mode. Be sure the correct memory bank is selected.
2. Press and hold down the SET button until three horizontal lines flash.
This indicates that all measurements have been erased, Fig. 3.
mmHg
M
D
AVG
(
( )
)
Fig. 2
Fig. 3